Image Engineering iQ-FoV Box Скачать руководство пользователя страница 4

 

   

 

 

 

 

 

Image Engineering 

 

Seite 

4

 von 

7

 

Please follow the steps below to 

correctly arrange your camera’s position and charts. If possible, use 

the live view mode of your camera. This mode will help you arrange the charts and align your camera. 

The magnetic 

rulers (see “image 03”) are an excellent way to align the camera properly, assuming that 

the camera’s orientation is correct when the horizontal/vertical rulers each show the same distance. 

3.1  Camera depth and height  

Attach the  camera  to  the mount  inside  the box  so  the back  door of  the box  can  be opened.  If  your 

camera is controllable via USB, you can use the cable already mounted inside the box and connect your 

control device from outside the box (the control panel at the front). The USB connection is placed right 

beside the illumination level knob. 

Align  the  camera  attachment  so  the  camera's  position  is  centered  inside  the  box  and  the  camera's 

optical axis is centered. For camera depth and height alignment, use the mechanical positioning of the 

extension arm from the outside of the box. Depending on your camera's FoV and dimensions, you may 

have to adjust the depth, height or leveling again when adjusting the chart positions, as described in the 

following chapters. We suggest starting with chart setup (1) to align the camera depth corresponding to 

the upper and lower chart in the image. These two charts should be placed inside the box as deep as 

possible, directly on the border of the back wall.  

3.2  Camera pitch and yaw  

The pitch and yaw alignment of your camera can be adjusted with the motorized positioner via the IR 

remote control from outside the box. The IR detector of the motor is positioned on the backside of the 

motor  device  on  a  small  pin.  You  can  adjust  this  pin  to  use  your  remote  control  device  from  other 

directions. 

3.3  Camera fine leveling  

The fine leveling of your camera can be regulated with the leveling tool  positioned directly under the 

camera attachment. 

 

 

Содержание iQ-FoV Box

Страница 1: ...Image Engineering GmbH Co KG Im Gleisdreieck 5 50169 Kerpen Germany T 49 2273 99 99 1 0 F 49 2273 99 99 1 10 www image engineering com iQ FoV Box User Manual August 8 2022 ...

Страница 2: ... SYSTEM 3 2 1 Placement 3 2 2 Starting Control overview 3 3 CAMERA AND CHART ALIGNMENT 3 3 1 Camera depth and height 4 3 2 Camera pitch and yaw 4 3 3 Camera fine leveling 4 3 4 Chart positioning 5 4 ILLUMINATION INTENSITY 6 5 CHART HANDLING 6 6 FIELD OF VIEW CALCULATION 7 ...

Страница 3: ...ox You need full access to the back door and enough space to adjust the alignment arm on the front side to the full length 2 2 Starting Control overview Connect the box to power and turn it on with the power switch placed on the control panel at the lower left edge of the front wall of the box On the right side of the control panel are the illuminance intensity knob the illumination on off button ...

Страница 4: ...ide the box and the camera s optical axis is centered For camera depth and height alignment use the mechanical positioning of the extension arm from the outside of the box Depending on your camera s FoV and dimensions you may have to adjust the depth height or leveling again when adjusting the chart positions as described in the following chapters We suggest starting with chart setup 1 to align th...

Страница 5: ...e chart C in the center of the back wall The other charts must be positioned on a straight vertical and horizontal line through the image center The vertical placement has to have one chart A on the top using the top wall and one chart B at the bottom using the bottom floor wall The horizontal placement has to be done similarly place one chart B on the left wall and one chart A on the right wall U...

Страница 6: ...l knob outside the device The intensity panel shows the actual intensity in percent Turn on off the tubes via the on off button To get the best performance measure adjust the intensity so that the brightest area of the OECF grey patch fields around the center chart C is on its possible maximum but not saturated You can put the density filters over the fluorescent tubes for low intensity use The ac...

Страница 7: ...distances a1 2 b1 2 and d in the following formula to calculate the angle of view for horizontal angle for vertical angle d is the distance the camera moved into the iQ FoV Box measured from the front plate see the left side of Image 04 The measure s origin is shown in Image 04 on the right side and is located in the center of the front plate Please note that it is essential to center the camera f...

Отзывы: