I.    TROUBLE SHOOTING 
 

 

 

Symptom 

Cause 

Remedy 

Part Code 

Coolant leaking between 
handpiece and connectors 

Defective O ring in 
Handpiece connector 

Replace O Ring 

O6890                 ( 
US O4231) 

Coolant leak between 
H‟piece and scaler tip 
insert 

Defective O ring in 
Handpiece 

Replace O Ring 

O6890   
(US O4232) 

3. 

Coolant in interior of 
Handpiece 

Defective O Ring in 
Handpiece 

Replace O Ring 

O6890   
(US O4233) 

4. 

Scaler tip vibrating, but no 
coolant spray 

Coolant line blocked.  
Flow control valve closed 

Check Filter 

– 

Open flow control 
valve 

W0530  
(US P7235) 
I6780  
(US P7056) 

Poor or no tip vibration 

Water in Handpiece 
 
 
Ferrite rod loose 
Broken Ferrite Rod 
Tip worn out 

Empty Handpiece 

– if 

re-occurs, replace 
Handpiece. 
Tighten Ferrite Rod 
Replace Ferrite Rod 
Replace tip 

 
 
O6900 (US O1002) 
See Pg 7 
O6870 (US O1200) 
See Pg 8 

6. 

No Power  

Blown fuse 

Replace Fuse Slow 
Blow 400mA 

– 250 v 

O6891 (US O2153) 

 

 
J.    CLEANING & STERILISING 

 
1.  Scaler tip insert (6), can be cleaned by chemical or steam autoclaving.  

Maximum 180º C/ 350ºF.  

 

2.  Handpiece (4).  

DO NOT Soak the Handpiece in liquid.

   

Clean by wiping 

with alcohol, soap or water based disinfectant solutions. 

 

DO NOT USE GLUTARALDEHYDE, FORMALIN SOLUTIONS

 or any 

products containing aggressive chemicals.  Do not put in ultrasonic cleaners, 
or soak in liquid chemicals. 

 

Steam sterilise the Handpiece at a maximum temperature of 121ºC / 250ºF 
(Rubber Pack Cycle). iM3 recommends autoclaving between patients. 
 
3.  Control Box & Cord (1) and (3) 
 
Wipe with alcohol, soap or water based disinfectant solution.  Solutions 
suitable for sanitising telephones are generally satisfactory. 
 

Содержание 42-12

Страница 1: ...es a truly unique instrument It not only scales in half the time of conventional Jack Hammer scalers but also produces less enamel damage due to its circular tip movement It can also scale sub gingiva...

Страница 2: ...2 42 12 Scaler Unit Components 2 1 5 4 3 17 6 7 8 9 10 13 18 12 15 11 16 14 9A see page 8...

Страница 3: ...needle valve I6780 US P7056 12 Coolant supply tube water or CLS maximum 3 bar O6930 13 Power Cord 110 240V AUS 06910 US O2197 14 Power Cord Socket power consumption 20 watts 15 Fuse Holder with spare...

Страница 4: ...e Handpiece accidentally falling on the floor this usually occurs when the Handpiece is placed on the table while adjusting the animal s head Position the Corian Handpiece Holder right at the animals...

Страница 5: ...t valve 11 is turned anti clockwise to allow the coolant to flow 3 Tip Vibration Power setting Amplitude Touching the power adjusting spot on the top of the Control Box 2 will alter the amplitude Powe...

Страница 6: ...s in a 360 circular motion easy access is guaranteed as all surfaces are active and there are no dead zones 5 The circular motion actually brushes the teeth rather than hammering them and produces mor...

Страница 7: ...eign matter is in the handpiece A loose ferrite rod is the most common cause of poor tip efficiency Change the ferrite rod to see if an improvement can be gained If no improvement is found a new tip i...

Страница 8: ...ANDLING CARE 1 The scaler tip inserts The shape of the tip is vital to the efficiency of the circular motion and ultrasonic effect DO NOT SHARPEN BEND OR RE SHAPE THE TIP Ferrite Rod Changing Instruct...

Страница 9: ...Perio Tip AUS O6850 US O0806 supplied with 42 12 Removes Supra and sub gingival calculi in all areas Particularly useful in sub gingival pockets as deep as 13mm The use of CLS will reduce the pathogen...

Страница 10: ...lace Handpiece Tighten Ferrite Rod Replace Ferrite Rod Replace tip O6900 US O1002 See Pg 7 O6870 US O1200 See Pg 8 6 No Power Blown fuse Replace Fuse Slow Blow 400mA 250 v O6891 US O2153 J CLEANING ST...

Страница 11: ...e 360 254 2981 Facsimile 360 254 2940 iM3 Pty Limited 9 31 33 Chaplin Drive LANE COVE NSW Australia 2066 Telephone 61 2 9420 5766 Facsimile 61 2 9420 5677 L TECHNICAL DATA Product Name iM3 42 12 Veter...

Страница 12: ...cifically created and designed by Philip Bloom of iM3 Australia exclusively for Veterinary Dental use No other ultrasonic dental scaler meets the specific needs of the Veterinary Dental Professional a...

Отзывы: