Make measurements with the sensor face at a 90-degree angle to the ceiling and floor and positioned in 

the center of the fixture (maximum UVC output). To test for proper UVC output, at a distance of 1 M from 

the center of the fixture raise the sensor into the UVC beam and find the maximum reading. For the 

Sanilume SL-36KT-75, the reading should be about 240 

𝜇𝜇𝜇𝜇

/

𝑐𝑐𝑚𝑚

2

 and 140 

𝜇𝜇𝜇𝜇

/

𝑐𝑐𝑚𝑚

2

 for the SL-36KT-40. 

To test for safety: 

• 

Measure every 3 ft. back from the fixture, up to 15 ft. 

• 

Measure near walls 

• 

Measure near any obstructions/protrusions that intersect with the UVC beam 

 

 

IF ANY READING IN THE ROOM IS ABOVE 0.2 µW/CM

2

 ANYWHERE ABOVE 7 FT. IN HEIGHT FROM 

THE FINISHED FLOOR, ADJUSTMENTS MUST BE MADE. 

 

1. Check the side-to-side level. If the apertures are not set the same or the power cable is 

pulling on the fixture, the level will be off. 

 

1.

 

If the reading that is too high is at the location where the UVC is being reflected down from 

the inside the fixture (about 3-7 ft from the fixture) First, try and raise the fixture by 1 or 2 

inches. Recheck the level and retest. 

 

2.

 

If the high readings are caused by ceiling reflections, then lower the aperture setting as 

previously described. 

 

 

 

Содержание SaniLume SL-36KT-40

Страница 1: ...ications Page 8 Certified Laboratory Test Results Page 9 Site Planning Installation Considerations Page 10 Site Planning Line of Sight Page 12 Site Planning Electrical Page 13 Site Planning F i x t u...

Страница 2: ...alls Since the reflective properties of surfaces can vary widely it should be considered as part of site planning Follow the manufacturer s recommendations for selecting appropriate ceiling and wall f...

Страница 3: ...th Danger indicates an IMMINENTLY hazardous situation which if not avoided WILL result in death or serious injury Warning indicates a POTENTIALLY hazardous situation which if not avoided COULD result...

Страница 4: ...es and warnings on the unit 15 SAVE THESE INSTRUCTIONS NOTICE These instructions do not claim to cover all details or variations in the equipment procedure or process described or to provide direction...

Страница 5: ...t the use of the Products will protect any user from or prevent infection and or contamination with any viruses bacteria protozoa fungi illness or disease The Products are not approved and or certifie...

Страница 6: ...ubmitted in writing within 30 days of discovery of the defect or failure to an authorized Illumisoft representative Product s or component part s may be required to be returned for inspection and veri...

Страница 7: ...ltra Quiet Fans Move 25 000 Cubic Feet Hr of Air Through The Fixture All Electronics Mounted In a Removable End Cap Features Light Shaping Baffle Vents Adjustable Aper tures For Different Ceiling Heig...

Страница 8: ...finished floor Recommended to not have obstructions or walls less than 10 ft in front of the fixture s apertures Compliance This GUV product is designed to comply with UL1598 CRD SL4 8 Risk Group 0 U...

Страница 9: ...9 SL OM ENG 16 09 2021 Wide Beam Narrow Beam Set for Higher Ceiling Heights Set For 8 9 Ft Ceilings...

Страница 10: ...10 SL OM ENG 16 09 2021 UVC Light UVC Light Top View Ceiling Floor 9 Ft Minimum Ceiling Height Sanilume UVC Distribution 7 9 Minimum...

Страница 11: ...A Sanilume with a 40 watt lamp can disinfect approximately 600 sq ft 55 7 sq m regardless of ceiling height Sanilume Orientation in Rooms As a general guideline the apertures that emit the UVC light s...

Страница 12: ...lings Glass UVC is blocked by glass so occupants in spaces with glass walls will be protected from UVC exposure Reflections UVC will reflect off walls and ceilings Ceiling reflections will be discusse...

Страница 13: ...d should use the SL 36KT 40 to avoid wall reflections that may exceed safety thresholds In general the higher the UVC output the greater the efficacy We recommend the 75 watt SL 36KT 75 version unless...

Страница 14: ...surements using a laboratory quality UV irradiance meter were taken with the readings being under 0 2 2 below 7 ft in height from the finished floor Important Note There are wide variations in reflect...

Страница 15: ...5 SL OM ENG 16 09 2021 Floor 7 ft Ceiling Wide Aperture Setting Un Safe Over 0 2 2 Floor 7 ft Ceiling Narrow Aperture Setting Safe Under 0 2 2 Fixture Mounting Heights Aperture Adjustment Site Plannin...

Страница 16: ...ral Rule 2 If the ceiling reflections are too high lower the aperture setting General Rule 3 If the reflections from the top louvers in the fixture are too high then raise the Sanilume closer to the c...

Страница 17: ...umes above any suspended light fixtures Ceiling Cut Away View With low ceiling heights 9 ft min and a high aperture setting the ceiling reflections will be relatively steep which will cause a shorter...

Страница 18: ...9 6 4 1 6 8 10 5 1 9 8 3 10 6 6 2 8 6 11 7 2 3 8 9 11 6 7 2 6 9 12 8 3 9 12 6 9 3 6 9 13 10 4 9 14 11 5 9 15 12 6 9 16 13 7 9 17 14 8 9 18 15 9 9 19 16 10 9 20 17 11 9 21 18 12 9 22 19 13 9 FOR CEILI...

Страница 19: ...C AREA 6 CEILING HEIGHT FT 9 FT MIN SQ FT QTY 40W QTY 75W 36KT 40 40 WATTS UVC LAMP 64 WATTS POWER 100 277 VAC 36KT 75 75 WATTS UVC LAMP 113 WATTS POWER 100 277 VAC AREA 7 CEILING HEIGHT FT 9 FT MIN S...

Страница 20: ...open during installation and maintenance 7 Denatured Alcohol Mounting Hardware Kit What s Included 1 2 x 1 16 Aircraft Cables 2 2 x T Bar Hangers 15 16 3 2 x 20 Bolts Nuts for Hangers 4 2 x 5 Canopies...

Страница 21: ...ld be hanging from the ceiling Thread one of the cables through one of the Griplock grippers and pull the end with a needle nose plyer towards the center of the fixture Do the same for the other end o...

Страница 22: ...s are installed newer tube designs may have increased irradiance Whenever modifications are made to the upper room GUV system or room e g adjustment of fixture height location or position of louvers a...

Страница 23: ...near walls Measure near any obstructions protrusions that intersect with the UVC beam IF ANY READING IN THE ROOM IS ABOVE 0 2 W CM2 ANYWHERE ABOVE 7 FT IN HEIGHT FROM THE FINISHED FLOOR ADJUSTMENTS M...

Страница 24: ...he Sanilume be cleaned at least once every six 6 months The Sanilume Germicidal Ultraviolet Lamp used in the Sanilume fixture has a manufacturer s rated average effective life of 10 000 hours Lamps ma...

Страница 25: ...o not get snagged on the clips 2 Verify that lamp is properly seated If not use a screwdriver at the end opposite of the connector 3 Close fixture to its operating position 4 Once the Sanilume is powe...

Страница 26: ...dusted from the bottom only Compressed air cans can be used as well Inspect the fans and dust them if needed Repeat the procedure with the opposing side of the Sanilume 3 Moisten a clean lint free clo...

Страница 27: ...maximum UVC output To test for proper UVC output at a distance of 1M from the center of the fixture raise the sensor into the UVC beam and find the maximum reading For the Sanilume SL 36KT 75 the rea...

Страница 28: ...18 patents and counting thistechnology enables their commercial light fixtures to achieve the highest energy efficiency in the world DLC Qualified Products Listings They are now applying their techni...

Отзывы: