6
Perform First Time Setup
4
Turn Toggle Switch and Press Power Button
5
Reset Both Default Passwords
1
Confirm Box Contents
For technical assistance, contact
Illumina Technical Support.
Website:
www.illumina.com
Email:
Telephone (North America):
+1.800.809.4566
Telephone (Outside North America):
+1.858.202.4566
3
Connect the System
2
Place on Lab Bench and Raise Monitor
iSEQ 100 REUSABLE TEST FLOW CELL
AND CARTRIDGE
Document # 1000000035963 v05
Material # 20024018
a
a
Turn the toggle switch on the back of the instrument to the power on (
I
) position.
b
When the power button on the left side of the instrument pulses, press it to turn on the control
computer and operating system.
Select Setup and follow the onscreen prompts to
perform a system check and configure the system.
For the system check, insert the reusable test flow
cell into the slot on the front of the reusable test
cartridge until it clicks. Make sure that the labels
face up.
After use, store at room temperature in the original
packaging.
b
a
To configure the iSeq 100 System for a
wired network connection, connect the
Ethernet cable to an Ethernet port.
b
Connect the power cord to the AC power
inlet on the rear panel, and then to a wall
outlet. An electrical ground is required.
a
b
The brown shipping box weighs 21 kg (47 lbs), and the instrument
weighs 16 kg (35 lbs). Both require a two-person lift.
Place the instrument on a level lab bench (see pictured minimum
clearances) with access to a power outlet. Lift the monitor from the top
of the instrument so that it stands upright.
Set aside the reusable test flow cell and cartridge for first time setup (step
6) and any future troubleshooting.
Store the reusable test flow cell and cartridge, spare drip tray pad, and
spare air filter in the original packaging at room temperature.
*Plug configuration might vary.
To prevent electrical shock do not remove
the outer panels from the instrument.
There are no user serviceable components
inside.
Visually inspect the power cord. The power
cord must be in good working condition
when using this device.
Only when the system is unplugged from
AC mains is all hazardous energy removed
from the system.
iSEQ 100 REUSABLE TEST
CARTRIDGE
iSEQ 100
REUSABLE
TEST FLOW
CELL
ETHERNET CABLE
POWER CORD*
iSEQ 100 SPARE
AIR FILTER
iSEQ 100
SPARE DRIP
TRAY PAD
This poster describes how to install
and set up the iSeq
TM
100 Sequencing
System.
For additional documentation,
training, safety information, and other
resources, visit:
Warnings
For Research Use Only. Not for use in diagnostic procedures.
© 2020 Illumina, Inc. All rights reserved. All trademarks are the property of Illumina, Inc. or their respective owners. For specific trademark information, see
www.illumina.com/company/legal.html
.
iSeq
100
Sequencing System
Setup Poster
support.illumina.com/iSeq100
a
Record the default password for
sbsadmin
that is displayed
on the orange screen. Log on with the default password, reset
the default password (minimum 10 characters), and continue to
follow the prompts.
b
Record the default password for
sbsuser
that is displayed
on the blue screen. Select the sbsuser account from the left
corner, and then log on with the default password. The status
bar glows green and the Setup screen appears.