ILFORD
SETTING PROCESSING TEMPERATURES
Press CLEAR, then 2 TEMP SET and follow the
instructions for setting the solution and dryer
temperatures, use the numbers on the keypad to
input data. Press ENTER, then CONFIRM then
CLEAR. The dryer minimum temperature is 45°C
(113°F).
The processing temperatures for ILFORD chemistry
can be checked on the control panel. Press 9 TEMP
DISPLAY to show the set and actual temperatures.
PROGRAMMING OPERATING TIMES
The date and time can be set using the control
panel, press CLEAR , then press 6 TIMER SET and
follow the instructions.
The processor can be programmed to automatically
re-start at a pre-set time so that the solutions reach
operating temperatures for the start of a day’s work.
Press CLEAR, then press 6 TIMER SET and follow the
instructions. To set the days you want the timer to
operate, press 0 for off, 1 for on.
DRAINING
Remove lower panel, withdraw bottle. Remove top
cap assembly. Remove drain cap and fit nozzle,
pour out the effluent. Refit caps, replace bottle and
refit panel. Each working tank can be drained by
opening the appropriate drain valve. Place a bottle
under the drain pipe (Dev or Fix) when draining
dev tank or fix tank. Water from the other tanks
goes to the main drain via the drain tray.
ADDING SOLUTIONS
Ensure the drain valves are closed. Unscrew and
withdraw the filter element assembly. Pour solution
into the temperature control tank as shown until
solution in the working tank overflows to the drain.
Refit the filter assembly.
Developer working tank capacity
- 10 litres
Fix working tank capacity
- 13 litres
Rinse solution working tank capacity - 4 litres
Each replenishment bottle holds 10 litres of solution,
top up as required.