• The unit is not suitable for manual operation. Observe all pertinent safety
instructions and directives, as well as work-safety and accident-prevention
regulations for laboratory use. The unit must not be operated without a
dispersion tool.
Never run dispersion tools dry, as the gasket and bearings will be
14
destroyed if the tools are not cooled by the medium. The gasket and
bearings are made of PTFE and rustproof steel; the following points should
therefore be noted:
Chemical reactions of PTFE occur in contact with molten or solute alkali
metals and alkaline earth metals, as well as with fine powders of metals in
groups 2 and 3 of the periodic system at temperatures above 300 °C - 400
°C. Only elementary fluorine, chlorotrifluoride and alkali metals attack it;
halogenated hydrocarbons have a reversible swelling effect.
(Source: Römpps Chemie-Lexikon and ”Ulmann”, Volume 19)
• Vibrations may loosen the hand screw on the flange of the drive unit. For
your own safety check that the hand screw is secure and tighten it if
necessary.
• Glass vessels must always be secured with a clamp to prevent them
spinning. When working with flasks elastic couplers must be used to
prevent fracture of the glass.
• Switch on the drive only at the lowest speed setting. Then gradually
increase the speed to the desired level.
• When working with the
R 50
agitator shaft the safety cage must not be
removed. The agitator shaft must be installed in the
T 50 basic
drive unit
such that the bent safety cage prevents any direct contact with the shaft
from in front.
• Only dispersion and agitating tools approved by IKA may be used!
• Even in the case of repair work, the unit must only be opened by a
qualified engineer. The mains plug must be pulled before opening the unit.
The voltage-carrying components inside the unit may continue to carry a
voltage for some time after the mains plug is pulled.
• Caution: Pull the mains plug before changing the dispersion heads!
• Caution: Continuous operation will cause the surface of the mounting
flange for the dispersion tool to become hot!
• The ventilation slots on the drive must not be obstructed!
• The device does not start up again after a current failure.
• The T 50 basic may not be used to process inflammable or explosive
media.
Contens
Safety instructions
082006