IKA RC 2 basic Скачать руководство пользователя страница 13

13

Set the first point temperature (step 1, e.g. 10 °C) with 

Pump (+)

” button (

F

) or “

Pump (-)

” button (

G

). Con-

firm the setting by pressing “

OK/Pump

“ button (

E

). 

Left area of the display appears the set value (10 °C) and 
the right area of the display appears the actual tempera-
ture value measured. The “

Set

” LED (

O

) flashes briefly. 

The unit now starts and controls to the set value. When 
the temperature has reached the set value and has be-
come constant, the “

Set

” LED (

O

) no longer flashes and 

the following display appears. The temperature mea-
sured by the unit appears on the left hand side.

Input the calibration value from the reference measuring 
instrument (e.g. 10.2 °C) with “

Pump (+)

” button (

F

) or

Pump (-)

” button (

G

).

Confirm the value by pressing the “

OK/Pump

“ button (

E

).

Back to previous screen for input a new value by press-
ing “

Temp

” button (

B

).

Back

OK

The first point calibration is finished now.
Calibration of the other points is performed in the same 
way.

Example: 2-point calibration:

Dip the temperature sensor of the reference measuring 
instrument into the bath fluid.

Select 2-point calibration in the menu. Press the “

OK/

Pump

“ button (

E

) to start the 2-point calibration.

 Interface and output

The device can be operated in “Remote” mode via the RS 
232 interface (

10

, see

Fig. 1

) or the USB interface (

11

, see

Fig. 1

) connected to a PC and with the laboratory soft-

ware Labworld

soft

®

.

Note: 

Please comply with the system requirements to-

gether with the operating instructions and help section 
included with the software.

USB interface:

The Universal Serial Bus (USB) is a serial bus for connect-
ing the device to the PC. Equipped with USB devices can 
be connected to a PC during operation (hot plugging). 
Connected devices and their properties are automatically 
recognized. The USB interface can also be used to update 
firmware.

USB device drivers:

First, download the latest driver for 

IKA

®

 devices with USB 

interface from 

http://www.ika.com/ika/lws/download/

usb-driver.zip

 and install the driver by running the setup 

file. Then connect the 

IKA

®

 device through the USB data 

cable to the PC. 
The data communication is via a virtual COM port. Con-
figuration, command syntax and commands of the virtual 
COM ports are as described in RS 232 interface.

RS 232 interface:

Configuration
- The functions of the interface connections between the 

stirrer machine and the automation system are chosen 
from the signals specified in EIA standard RS 232 in ac-
cordance with DIN 66 020 Part 1.

-  For the electrical characteristics of the interface and the 

allocation of signal status, standard RS 232 applies in ac-
cordance with DIN 66 259 Part 1.

-  Transmission  procedure:  asynchronous  character  trans-

mission in start-stop mode.

- Type of transmission: full duplex.
- Character format: character representation in accordance 

with data format in DIN 66 022 for start-stop mode. 1 
start bit; 7 character bits; 1 parity bit (even); 1 stop bit.

- Transmission speed: 9600 bit/s.
- Data flow control: none
- Access procedure: data transfer from the stirrer machine to 

the computer takes place only at the computer’s request.

Содержание RC 2 basic

Страница 1: ...20000004342 IKA RC 2 basic RC 2_112014 Reg No 4343 01 Operating instructions EN...

Страница 2: ...7 6 8 10 11 12 12 4 9 Item Designation 1 Operator panel and display 2 Mains switch 3 Power socket 4 Venting grid 5 Filling lid 6 Pump connection IN 7 Pump connection OUT 8 Overflow 9 Backflow 10 RS 2...

Страница 3: ...ation for IKA fluid 8 Moving the device 9 Operator panel and display 9 Commissioning 10 Menu settings 11 Menu structure 11 Basic guide to setting the menu options 11 Operating mode MODE 11 Maximum tem...

Страница 4: ...the device may still be live for some time after unplugging from the power supply Coverings or parts that can be re moved from the device must be put back on the device again to ensure safe operation...

Страница 5: ...re of the chemical hazards that may be associated with the bath fluid used Observe all safety warning for the fluids Depending on the bath fluid used and the type of opera tion or toxic can arise Ensu...

Страница 6: ...peration Connecting the tubings Unscrew the union nuts and stoppers using a wrench from the pump connection IN 6 see Fig 1 and OUT 7 see Fig 1 External system IKA Calorimeter Fig 5 Connect the hoses f...

Страница 7: ...rning message appear on the display Meanwhile the bottom LED segment of the lights up in red color to indicate the low fluid level see Fig 8 Open the filling lid 5 see Fig 1 and fill fluid to the bath...

Страница 8: ...High purity water from icon exchangers and distilled or bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a m...

Страница 9: ...se the temperature setting Navigation selecting the settings in the menu D Temp button Decrease the temperature setting Navigation selecting the settings in the menu E OK Pump button Start stop the pu...

Страница 10: ...ng with Temp button C or Temp button D Change the pump speed setting with Pump button F or Pump button G In standby status activate the refrigerating function by pressing the Temp button B The pump an...

Страница 11: ...nu by pressing Menu key H F Press the Temp button C or Temp button D to change the menu options F Press the Pump button F or Pump button G to change the menu value settings F Confirm the menu settings...

Страница 12: ...isting control deviation A high Ti means a smaller and slower effect on the control variable Ti values that are too small can lead to instability of the controller The differential time Td s is the d...

Страница 13: ...ting instructions and help section included with the software USB interface The Universal Serial Bus USB is a serial bus for connect ing the device to the PC Equipped with USB devices can be connected...

Страница 14: ...MS Windows and includes graphical entry features for pump motor speed ramps for example PC 1 1 Cable This cable is required to connect RS 232 port 10 to a PC 1 2 RxD 3TxD 4 5 GND 6 7 RTS 8 CTS 9 1 Rx...

Страница 15: ...ing Reduce pump motor speed Use fluid with lower viscosity Check if the pump impeller is blocked Err 06 Pump off Refrigerating off Too low liquid level Check the fluid level and buoyage Err 15 Pump of...

Страница 16: ...rding to DIN 12876 1 Class III FL suitable for flammable and non flammable fluids Cooling capacity according to DIN 12876 20 C 10 C 0 C 10 C 20 C W 400 370 320 240 130 Pump speed adjustable rpm 1000 3...

Страница 17: ...iv ing reasons for the claim You will be liable for freight costs The warranty does not cover worn out parts nor does it apply to faults resulting from improper use insufficient care or maintenance no...

Страница 18: ...ng the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 1245 600560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: