8
Delivery scope:
Telescopic stand
-
R 472 Floor stand
or
R 474 Telescopic stand or T 653
Tele-
scopic stand
according to order
- Operating instructions.
For technical reasons associated with shipping, the telescopic stand is
supplied stripped down, as pre-assembled modules.
Set-up and commissioning
Intended use
Use:
The
R 472 Floor stand / R 474
Telescopic stand
is used solely for
mounting
RW 47 digital
agitator units.
WARNING
R 472 Floor stand:
the castors must be locked when the device is in use.
The
T 653 Telescopic stand
is used solely for mounting
T 65 basic
/ digital ULTRA-TURRAX
®
dispersion units.
•
The handwheel bolt on the stand can come loose due to vibration,
which would lead to the unit moving down. So, check the tightness
of the handwheel bolt regularly when using the unit, and retighten
it if necessary.
•
Raise or lower the stand (after slackening the handwheel bolt) by
applying the manual force as close as possible to the stand pillar.
The further the force is applied from the pillar, the more force you
need to apply, plus there is an increased risk of the equipment top-
pling over.
Unpacking
Unpacking:
- Please unpack the device carefully.
- In the case of any damage a detailed report must be sent immedi-
ately (post, rail or forwarder).
The safety of you cannot be guaranteed:
- If the device is operated with accessories that are not supplied or
recommended by
IKA
.
- If the device is operated improperly or in contrary to the
IKA
spec-
ifications.
NOTICE
• Secure the vessel against spinning round, by using a clamp, for ex-
ample.
• Observe the operating instructions of the
T 65 basic / digital UL-
TRA-TURRAX
®
dispersing unit or the
RW 47 digital
stirring unit!
• Check the appliance and accessories beforehand for damage each
time you use them. Do not use damaged components.
• Protect the appliance and accessories from bumps and impacts.
Setting up the stand (R 472 Floor stand):
Place
RW 47 digital
overhead stirrer on the arm of the stand. Fasten the
overhead stirrer to the stand from below with hexagonal socket screws sup-
plied. The help of an assistant is needed for screwing on the overhead stirrer.
The cable clips included in the delivery package are required for fixing the
power cable of the unit.
Check that the stirrer is held in position firmly prior to each use and also at
regular intervals. The position of the stirrer must only be adjusted when the
equipment is stationary and the power supply is disconnected.
Содержание R 472
Страница 15: ......