IKA MultiDrive basic Скачать руководство пользователя страница 4

6

7

Safety instructions

/// General information

› 

Read the operating instructions completely before starting up and follow the safety 

instructions.

›  Keep the operating instructions in a place where it can be accessed by everyone�
›  Ensure that only trained staff work with the device�
›  Follow the safety instructions, guidelines, occupational health and safety and accident 

prevention regulations�

›  The device must only be used in a technically perfect condition�

 Caution!   

›  Wear your personal protective equipment in accordance with the hazard category of the 

medium to be processed, there is a risk of:
- splashing of liquids
- body parts, hair, clothing and jewellery getting caught
- sharp blade edges�

!

 Notice!

›  Pay attention to the marked sites in Fig� 1� 

/// Device setup

 Caution!

›  The power switch of the device must be accessed immediately, directly and without risk at 

any time� If access to the power switch cannot be ensured, an additional emergency stop 
switch that can be easily accessed must be installed in the work area�

!

 Notice!

›  Set up the device in a spacious area on an even, stable, clean, non-slip, dry and fireproof surface�
›  Make sure that the ventilation slits are not blocked at back of the device�
›  Ensure that the vessel and the cover is firmly attached to the drive unit prior to operating the 

device�

›  The drive unit cannot be operated without a closed vessel� 

EU Declaration of conformity

We declare under our sole responsibility that this product corresponds to the directives 2014/35/EU, 
2006/42/EC, 2014/30/EU and 2011/65EUand conforms with the following standards or normative 
documents: EN 61010-1, EN 61010-2-051, EN ISO 12100, EN 61326-1 and EN 60529�
A copy of the complete EU Declaration of conformity can be requested at sales@ika�com�

Source language: German

Explication of warning symbols

/// Warning symbols

Indicates an (extremely) hazardous situation, which, if not avoided, will 
result in death, serious injury�

Danger!

Indicates a hazardous situation, which, if not avoided, can result in 
death, serious injury�

Warning!

Indicates a potentially hazardous situation, which, if not avoided, can 
result in injury�

Caution!

Indicates practices which, if not avoided, can result in equipment damage�

Notice!

/// General Symbols

Position number

A–––

Correct/Result
Shows the correct execution or the result of an action step�

Wrong
Shows the incorrect execution of an action step� 

Note
Displays action steps that require particular attention to detail� 

/// Working with the device

  Danger!  

›  Do not use the device in explosive atmospheres, it is not EX-protected� 
›  With substances capable of forming an explosive mixture, appropriate safety measures must 

be applied, e�g� working under a fume hood�

›  To avoid body injury and property damage, observe the relevant safety and accident preven-

tion measures when processing hazardous materials�

 

Warning!    

›  Only process samples that will not react dangerously to the extra energy produced through 

processing� This also applies to any extra energy produced in other ways, e�g� through light 
irradiation�

!

!

Содержание MultiDrive basic

Страница 1: ...IKA MultiDrive basic...

Страница 2: ...3 ENGLISH...

Страница 3: ...ion of warning symbols 6 Safety instructions 7 Unpacking 12 Operation 16 Error codes 21 Intended use 9 Technical Data 25 Interfaces and outputs 23 Warranty 26 Maintenance and cleaning 22 Accessories 2...

Страница 4: ...r sole responsibility that this product corresponds to the directives 2014 35 EU 2006 42 EC 2014 30 EU and 2011 65EUand conforms with the following standards or normative documents EN 61010 1 EN 61010...

Страница 5: ...erial leaks from the vessel Clean the device after disconnected power plug The feet of the device must be clean and undamaged The device is designed for short term operation The maximum operating time...

Страница 6: ...iron metals The feed material must not be too moist or fatty Otherwise the material may adhere to the sides of the grinding container Following is a list of some materials that can be reduced Leaves...

Страница 7: ...indicator Indicate the interval function is activated B Timer indicator Indicate the timer function is activated C Counter indicator Indicate the counter function is activated D Rotating pressing knob...

Страница 8: ...the drive unit 3 2000 ml max 1500 ml 500 ml min 1000 ml Filling the vessel Closing the vessel Notice The socket used must be earthed fitted with earth contact Observe the ambient conditions temperatur...

Страница 9: ...ing the speed 4 Interval Timer Counter Speed x 1000 rpm min sec Interval Timer Counter Speed x 1000 rpm min sec Interval Timer Counter Speed x 1000 rpm min sec green Note After the cover has been secu...

Страница 10: ...Timer editing LED Interval Timer Counter min sec Interval Timer Counter min sec Counter Timer Interval Timer Counter min sec Note If the counter function is activated the counter automatically starts...

Страница 11: ...r including a short description of the fault Causes Inside temperature is too hot Effect motor off Solutions Switch off the device and allow it to cool down Restart the instrument Causes Motor can t r...

Страница 12: ...erfaces and outputs The device can be connected to a PC and operated with the laboratory software labworldsoft through USB interface Note Please comply with the system requirements together with the o...

Страница 13: ...to xxx START_4 Start the motor STOP_4 Stop the motor RESET Switch to normal operation mode Accessories IKA MultiDrive BL 2000 blending vessel IKA MultiDrive MI 250 milling vessel IKA MultiDrive MI 400...

Страница 14: ...ay also send the machine direct to our factory enclosing the delivery invoice and giving reasons for the claim You will be liable for freight costs The warranty does not cover worn out parts nor does...

Страница 15: ...and LTD Phone 44 1865 986 162 eMail sales england ika com Malaysia IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn...