Interface and output

The device can be connected to a PC and operated with the 
laboratory software  labworld

soft

®  

through the RS 232 port, 

USB port or USB port on the 

WiCo

Note: 

Please observe the system requirements together 

with the operating instructions and help section included 
with the software�

USB interface:

The Universal Serial Bus (USB) is a serial bus for connecting 
the device to the PC� Equipped with USB devices can be 
connected to a PC during operation (hot plugging)� Con-
nected devices and their properties are automatically recog-
nized� Use the USB port in conjunction with labworldsoft

®

 

for operation in “Remote” mode and also to update the 
firmware�

USB device drivers:

First, download the latest driver for 

IKA

®

 devices with USB 

port from: 

http://www.ika.com/ika/lws/download/usb-driver.zip.

Install the driver by running the setup file� Then connect the 

IKA

®

 device through the USB data cable to the PC� 

The data communication is via a virtual COM port� Configu-
ration, command syntax and commands of the virtual COM 
ports are as described in RS 232 port�

Device software update:

For device software update, visit 

IKA

®

 website 

www.ika.

com

 and enter the “Service” menu� Download and run the 

Firmware Update Tool�
Find and click the Firmware Update Tool in your PC after 
installation� Register your E-mail and password� 
Connect the device to your PC via the USB cable� Then, you 
can update the device software according to instructions of 
the Firmware Update Tool�

RS 232 interface:

Configuration
- The functions of the interface connections between the 

stirrer machine and the automation system are chosen 
from the signals specified in EIA standard RS 232 in ac-
cordance with DIN 66 020 Part 1�

- For the electrical characteristics of the interface and the al-

location of signal status, standard RS 232 applies in accor-
dance with DIN 66 259 Part 1�

- Transmission procedure: asynchronous character transmis-

sion in start-stop mode�

- Type of transmission: full duplex�
-  Character format: character representation in accordance 

with data format in DIN 66 022 for start-stop mode� 1 
start bit; 7 character bits; 1 parity bit (even); 1 stop bit�

- Transmission speed: 9600 bit/s�
- Data flow control: none
- Access procedure: data transfer from the stirrer machine to 

the computer takes place only at the computer’s request�

Command syntax and format:

The following applies to the command set:
- Commands are generally sent from the computer (Master) 

to the device (Slave)�

- The device sends only at the computer’s request� Even 

fault indications cannot be sent spontaneously from the 
device to the computer (automation system)�

-  Commands are transmitted in capital letters�
- Commands and parameters including successive param-

eters are separated by at least one space (Code: hex 0x20)�

-  Each individual command (incl� parameters and data) and 

each response are terminated with CR LF (Code: hex 0x0d 
hex 0x0A) and have a maximum length of 80 characters�

- The decimal separator in a number is a dot (Code: hex 

0x2E)�

The above details correspond as far as possible to the recommendations of the NAMUR working party (NAMUR recommenda-
tions for the design of electrical plug connections for analogue and digital signal transmission on individual items of laboratory 
control equipment, rev� 1�1)�
The NAMUR commands and the additional specific

 IKA

®

 co mmands serve only as low level commands for communication 

between the device and the PC� With a suitable terminal or communications programme these commands can be transmitted 
directly to the circulator equipment� The 

IKA

®

 software package, Labworld

soft

®

, provides a convenient tool for controlling 

circulating equipment and collecting data under MS Windows, and includes graphical entry features, for pump motor speed 
ramps for example�

Commands
NAMUR Commands Function

IN_PV_1

Read the external actual temperature

IN_PV_2

Read the internal actual temperature

IN_PV_3

Read the safety actual temperature

IN_PV_4

Read the pump actual speed

IN_SP_1

Read the internal setting temperature (if 0: internal control)
Read the external setting temperature (if 1: external control)

IN_SP_3

Read the safety setting temperature

Содержание HRC 2 control

Страница 1: ...Operating instructions EN 20000015725a IKA HRC 2 control HRC 2 control_112017...

Страница 2: ...Designation 1 Wireless Controller WiCo 2 Handle 3 Venting grid 4 Filling opening lid 5 RS 232 port 6 USB port 7 External temperature sensor socket 8 Multifunction port 9 Overflow 10 Backflow 11 Power...

Страница 3: ...he WiCo 9 Mounting the WiCo to the station 9 Filling and draining 9 Fluid Standard information for IKA fluid 11 Moving the device 12 Charging the Battery Pack RB 1 rechargeable battery 12 Changing bat...

Страница 4: ...rates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equip...

Страница 5: ...inhibit fire or ex plosions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please contact IKA application sup port if you have any...

Страница 6: ...union nuts caps of the pump outputs and inputs and blow compressed air through all the pump outputs and inputs Beware of the risk of burning due to delay in boiling Fluids Beware of the risk of burnin...

Страница 7: ...below 20 C Only the rechargeable battery types recommended in the technical data may be used in the device Do not charge batteries that have leaked or that are discolored deformed or damaged in any ot...

Страница 8: ...nection M 16 x 1 for IN and OUT directly or with the olives Screw the hose olives to the pump connection IN and OUT with union nuts Slide the hoses NW 12 onto the olives The hoses must be secured with...

Страница 9: ...S IP 40 Model WiCo CBC 5 LPS IP 40 Note notch upward Mounting the WiCo to the station Note If the WiCo need be permanently attached to the station we strictly recommend to fasten the unlocking button...

Страница 10: ...rain valve in counter clockwise direction with a straight screwdriver Fluid level information on the fluid level indicator Red Low Level Red High Level Fig 12 Note Pay attention to the fluid level inf...

Страница 11: ...e Temperature 4 Viscosity Viscosity at 25 C for Heating Fluid HF Viscosity at 20 C for Cooling Fluid CF Viscosity at 25 C for Universal Fluid UF Dynamic viscosity mPa s is a product of kinematic visco...

Страница 12: ...oss head clamp H 38 Holding rod IKA HRC 2control IKA HRC 2 control IKA HRC 2control Moving the device Empty all fluid in the bath before moving device from one place to other place The device can be l...

Страница 13: ...evel U Menu button Press it once main menu is displayed Press it a second time back to the working screen V Lock button Lock Unlock the knobs and buttons Item Designation Function A On Off button Swit...

Страница 14: ...ssing the power switch 12 of the station the screen of the WiCo displays the device designation and the software version after a beep IKA Circulator HRC Control Version 1 2 008 Fig 20 After several se...

Страница 15: ...station the USB icon appears on the screen If the WiCo is not connected with station via a USB cable the data exchange between the station and the WiCo via Bluetooth In this case the Bluetooth icon i...

Страница 16: ...This symbol indicates that the pump is activated Warning This symbol indicates that warning is active PC PC Control This symbol means that either the station or the WiCo is connected to a computer and...

Страница 17: ...f you press the Menu button U the system skips directly back to the working screen If you press the Back button T the system skips back to the previous screen Menu Menu navigation Press the Menu butto...

Страница 18: ...nment Set Temperature Actual Internal Temp Actual External Temp X Axis Y Axis 15 min 30 min 60 min 120 min Axis Scaling Maximum Minimum activated activated activated activated activated 85 C 5 C PROGR...

Страница 19: ...be used in the event of special tem perature control requirements When Manual is selected the following parameters can be set for Internal int and External ext temperature control Kp Proportional coef...

Страница 20: ...2 OK Calibration Internal 2 Point Calibration Step 1 2 Enter the temperature from the measuring device 5 00 OC OK Calibration Internal 2 Point Calibration Step 1 2 Enter the temperature from the meas...

Страница 21: ...The scaling of the temperature axis Y axis is automati cally determined depending on the target temperature and the actual internal and external temperature The time axis X axis is permanently scaled...

Страница 22: ...ult Status Define the default state OFF of the external valve as Open or Closed It depends on the valve type nor mally open or closed 3 Refill Use the external M1 valve to automatically check the leve...

Страница 23: ...specified temperature Note The factory setting of the safe speed is 2000 rpm and is activated after the time limit is set see Set 3 Temperature In the Temperature menu option you can specify a tempera...

Страница 24: ...ll duplex Character format character representation in accordance with data format in DIN 66 022 for start stop mode 1 start bit 7 character bits 1 parity bit even 1 stop bit Transmission speed 9600 b...

Страница 25: ...ature control OUT_WD1 n Start the watchdog mode 1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the tempering and pump functions are switched off...

Страница 26: ...y 0V only for Standby 7 reserved for later use do not connect 8 reserved for later use do not connect 1 2 3 4 5 6 7 8 Fig 35 Multifunction port USB micro B USB micro B USB micro A USB micro A USB cabl...

Страница 27: ...e isopropyl alcohol Foodstuffs Water containing tenside Fuel Water containing tenside For materials which are not listed please request information from IKA application support Wear protective gloves...

Страница 28: ...eating refrigerating off Heater switched off by safety circuit Check safety temperature set value fluid level Error 15 Pump off Heating refrigerating off Safety shutdown of the compressor Restart the...

Страница 29: ...licone tube nominal width 12 H PUR 8 PUR tube nominal width 8 mm H PUR 12 PUR tube nominal width 12 mm H FKM 8 FKM tube nominal width 8 mm H FKM 12 FKM tube nominal width 12 mm Tubing insulations ISO...

Страница 30: ...olution K 0 01 Classification according to DIN 12876 1 Class III FL suitable for flammable and non flammable fluids Safety circuit adjustable C 0 110 Safety temperature display TFT LCD on WiCo Heating...

Страница 31: ...ient temperature C 5 40 Ambient humidity relative 80 IP code according to EN 60529 IP 40 Interface USB RB 1 Battery pack Voltage V 3 7 Charging capacity mAh 2000 Charging time h 4 5 Working time h 15...

Страница 32: ...ble for freight costs The warranty does not cover worn out parts nor does it apply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with the in stru...

Страница 33: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Отзывы: