PT3 device is intended for use in an electromagnetic environment 

in which radiated RF disturbances are controlled. The customer or 

the user of the PT3 device can help prevent electromagnetic 

interference by maintaining a minimum distance between 

portable and mobile RF communications equipment (transmitters) 

and the PT3 as recommended below, according to the maximum 

output power of the communication sequipment.

Rated maximum

output power of

transmitter W

150kHz to 

80MHz

80MHz to 

800MHz

800MHz to 

2,5GHz

Recommended separation distances between portable and 

mobile RF communications equipment and the PT3 device.

0.01

0.1

1

10

100

0.12

0.38

1.2

7.8

12

0.12

0.38

1.2

7.8

12

0.23

0.73

2.3

3.3

23

d=1.2   p

d=1.2   p

d=2.3   p

Separation distance according to frequency 

of transmitter m

Electrostatic

discharge(ESD)

lEC61000-4-2

WARRANTY

Please contact your dealer or the device center in case of a 
claim under the warranty. If you have to send in the unit, 
enclose a copy of your receipt with clear statement of defect 
description.
The warranty terms as below:

1. The warranty period for device is one year from date of 

delivery. In case of a warranty claim, the date of delivery 
has to be proven by means of the sales receipt or invoice.

2. Repairs under warranty do not extend the warranty period.
3. The following cases are excluded under the warranty

4. Liability for direct or indirect consequential losses 

caused by the unit is excluded even if the damage to the 
unit is accepted as a warranty claim.

     All damage which has arisen due to improper 
treatment, e.g. nonobservance of the user instruction.
     All damage which is due to repairs or tampering by the 
customer or unauthorized third parities.
     Damage which has arisen during transport from the 
manufacturer to the consumer or during transport to the 
service centre.
     Accessories which are subject to normal wear and tear.

Manufactured for iHealth Labs, Inc.
USA:
iHealth Labs, Inc. 
www.ihealthlabs.com  
120 San Lucar Ct., Sunnyvale, CA 94086, USA
+1-855-816-7705 

Famidoc Technology Co., Ltd.
No. 212 Yilong Road, Hexi Industrial Zone, 
Jinxia, Changan Town, Dongguan 523853, 
Guangdong Province, China 

Europe:
iHealthLabs Europe SAS
www.ihealthlabs.eu  
36 rue de Ponthieu, 75008, Paris, France 

iHealthLabs Europe SAS
www.ihealthlabs.eu  
36 rue de Ponthieu, 75008, Paris, France 

0120

Immunity test

Compliance 

level

ConductedRF

lEC6 1000-4-6

Radiated RF lEC

61000-4-3

3V rms

150 kHzto

80 MHz 

3V rms

3V/m 

3V/m

80 MHz to 

2.5 GHz

±6kV 

contact

±8k Vair

Portable and mobile RF 

communications 

equipment should be 

used no closer to any 

part of the PT3 device, 

than the recommended 

separation distance 

calculated from the 

equation applicable to 

the frequency of the transmitter.

Recommended separation distance.

where P is the maximum output power rating of the transmitter in 

watts (W) according to the. transmitter manufacturer and d is the 

recommended separation distance in meters(m). Field strengths 

from fixed RF transmitters, as determined by an electromagnetic 

site survey, a. should be less than the compliance level in each 

frequency range.

b. Interference may occur in the vicinity of equipment 

marked with the following symbol:

NOTE I  At 80 MHz end 800 MHz. the higher frequency range 

applies.

NOTE2 These guidelines may not apply in all situations. 

Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

a. Field strengths from fixed transmitters, such as base stations 

for radio(cellular/cordless) telephones and land mobile radios, 

amateur radio, AM and FM radio broadcast and TV broadcast 

cannot be predicted theoretically with accuracy. To assess the 

electromagnetic environment due to fixed RF transmitters, an 

electromagnetic site survey should be considered. If the 

measured field strength in the location in which the PT3 

device is used exceeds the applicable RF compliance level 

above, the should be observed to verify normal operation. If 

abnormal performance is observed, additional measures may 

be necessary, such as reorienting or relocating the PT3.

b. Over the frequency range 150 kHz to 80 MHz, field strengths 

should be less than [Vi] V/m.

IEC 60501 

Test level

Electromagnetic 

environment - guidance

d=1.2   p
d=1.2   p 80MHz

800MHz

d=1.2   p 800MHz

2.5GHz

For transmitters rated at a maximum output power not listed 

above, recommended separation distance d in meters(m) can 

be estimated using the equation applicable to the frequency of 

the transmitter, where P is the maximum output power rating of 

the transmitter in watts(W) accordable to the transmitter 

manufacturer.

NOTEI At 80 MHz and 800 MHz. the separation distance for the 

higher frequency range applies.

NOTE2 These guidelines may not apply in all situations. 

Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

10

11

Содержание PT3

Страница 1: ...User s Manual Mode d emploi Manual de usuario Infrared No Touch Forehead Thermometer l infrarouge Pas de contact forhead thermomètre Infrarrojos Termómetro sin contacto frente Version 1 0 ...

Страница 2: ...body 7 If there is a temperature difference between the thermometer storage area and the new ambient environment around the subject please let the thermometer sit within the new environment for 30 minutes before taking the measurement 8 Do not measure body temperature immediately after consuming a drug that raises body temperature Temperature readings taken at this time will not be accurate 9 It i...

Страница 3: ...tandards test no toxicity allergy and irritation reaction They are in compliance with the MDD requirements Based on the current science and technology other potential allergic reactions are unknown 2 About the product measurement accuracy We recommend taking only up to 3 continuous readings within a short period 10 Do not directly face the sun or an air outlet of an air conditioning or radiator de...

Страница 4: ... during installation Make sure that the new batteries are tightly inserted into the battery compartment and make sure that the polarity is not reversed during installation 4 Return the battery cover to close the battery compartment Comply with relevant national laws and regulations when disposing of the used batteries Do not dispose of batteries directly into the trash bag Remove the batteries if ...

Страница 5: ...orities responsible for waste disposal Notes Please act according to local laws for disposal of used batteries Take out the battery if you are not going to use the unit for a long time Body temperature varies from person to person and fluctuates during the course of the day For this reason it is suggested to know one s normal healthy forehead temperature to correctly determine the temperature Body...

Страница 6: ...n environment Immunity test IEC 60601 Test level 6kV contact 8kV air Compliance level 6kV contact 8kV air Attention Follow instructions for use The sign background color blue The sign graphical symbol white Authorized representative in the European Community Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Elec...

Страница 7: ...e 0120 Immunity test Compliance level ConductedRF lEC6 1000 4 6 Radiated RF lEC 61000 4 3 3V rms 150 kHzto 80 MHz 3V rms 3V m 3V m 80 MHz to 2 5 GHz 6kV contact 8k Vair Portable and mobile RF communications equipment should be used no closer to any part of the PT3 device than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended...

Страница 8: ...tes ou des écorchures 6 Ne mesurez pas la température corporelle au desuss de pansements situés sur le front car ils affectent la conduction thermique 7 S il y a une différence de température entre le lieu de rangement du thermomètre et l environnement où se trouve le sujet laissez le thermomètre au repos pendant 30 minutes avant d effectuer une mesure 8 Ne mesurez pas la température immédiatement...

Страница 9: ... 10993 5 et ISO 10993 10 aucune toxicité allergie et réaction 2 À propos du produit continues la température corporelle du patient peut se transmettre au thermomètre et affecter la précision des mesures Nous vous recommandons d effectuer un maximum de 3 mesures continues sur un court laps de temps 10 Ne pointez pas directement l appareil vers le soleil une climatisation ou un chauffage pendant la ...

Страница 10: ...a Rappel Si vous n appuyez pas longuement 8 secondes sur le bouton de mesure pour confirmer l unité sélectionnée avant que l écran ne s éteigne le système considérera que la modification a échoué et retiendra l unité de température d origine 1 Si l appareil est éteint appuyez longuement 8 secondes sur le bouton de mesure La mode de selection d unité s affiche Dans ce mode les voyants C et F cligno...

Страница 11: ...rature détectée avec la température normale de la personne Une élévation de la température corporelle de référence de 1 F 1 C ou plus est généralement une indication de fièvre Différents sites de mesure rectale axillaire orale frontale auriculaire donneront des lectures différentes Il est donc faux de comparer les mesures prises sur différents sites Voici les températures typiques pour les adultes...

Страница 12: ... ses fonctions Les appareils électroniques situés à proximité peuvent s en voir affectés Émissions RF CISPR11 Émissions harmoniques CEI 61000 3 2 Émissions par fluctuations ou papillotement de la tension CEI 61000 3 3 Classe B Sans objet Sans objet Le PT3 peut être utilisé dans toute installation autre que résidentielle ou directement connectées au réseau public basse tension desservant des immeub...

Страница 13: ...étique Le PT3 est conçu pour être utilisé dans un environnement électromagnétique tel que spécifié ci après Le client ou l utilisateur du PT3 doit assurer qu il soit utilisé dans un tel environnement Test d immunité Niveau de conformité RF conduite CEI 61000 4 6 RF rayonnée CEI 61000 4 3 3 V rms 150 kHz à 80 MHz 3 V rms 3 V m 3 V m 80 MHz à 2 5 GHz 6 kV contact 8 kV air Les appareils de communicat...

Страница 14: ...ctes causées par l appareil est exclue même si la réclamation pour des dommages de l appareil est acceptée au titre de la garantie Tout dommage découlant d un traitement inapproprié par exemple du non respect des instructions Tout dommage découlant de réparations ou d interventions faites par le client ou par des tiers non agréés Tout dommage survenu lors du transport du fabricant au consommateur ...

Страница 15: ...as o abrasiones 6 No mida la temperatura corporal de tejidos cicatrizados en la frente ya que este tipo de tejidos alteran conducción de la temperatura del cuerpo 7 Si existe una diferencia de temperatura entre el sitio donde guarda el termómetro y el nuevo entorno alrededor del sujeto deje que el termómetro se asiente en el nuevo entorno durante 30 minutos antes de tomar la medida 8 No mida la te...

Страница 16: ...s aplicaciones Contraindicaciones Este producto utiliza principalmente un detector de temperatura por IR para medir la frente y determinar la temperatura corporal humana Se puede usar en bebés niños y adultos Los bebés y los niños no deben utilizar el termómetro por sí mismos Las lecturas de temperaturas corporales de bebés y niños deben ser tomadas por un adulto Aviso Las lecturas de temperatura ...

Страница 17: ...eratura del entorno no cumple los requisitos de medición o hay una diferencia de temperatura excesivamente grande C La temperatura objetivo excede del intervalo de medición Errores del producto y solución de problemas Elemento a detectar Solución Problema Pila agotada Las pilas no se han instalado adecuadamente No se puede realizar la medida ya que la capacidad actual de la pila es demasiado baja ...

Страница 18: ...cionamiento del dispositivo 2 El termómetro IR es un instrumento extremadamente preciso Cualquier mantenimiento inadecuado Temperatura corporal La temperatura corporal varía de persona a persona y fluctúa durante el transcurso del día Por esta razón se sugiere conocer la temperatura normal y saludable de la frente para determinar correctamente la temperatura La temperatura corporal corre aproximad...

Страница 19: ...e del dispositivo PT3 que la distancia de separación recomendada calculada a partir de la ecuación aplicable a la frecuencia del transmisor Distancia de separación recomendada donde P es la potencia de salida máxima asignada al transmisor en vatios W de acuerdo con el fabricante del transmisor y d es la distancia de separación recomendada en metros m La intensidad de campo procedente de transmisor...

Страница 20: ...timarse utilizando la ecuación aplicable a la frecuencia del transmisor donde P es la potencia de salida máxima asignada al transmisor en vatios W de acuerdo con el fabricante del transmisor NOTA 1 A 80 MHz y 800 MHz se aplica la distancia de separación para la gama de frecuencias más alta NOTA 2 Es posible que estas directrices no sean de aplicación en todas las situaciones La propagación electro...

Отзывы: