    1) The coefficient of variation (CV) of pulse period >25%.

    2) The difference of adjacent pulse period is ≥0.14s and more than 53 percent of the total 

        number of pulses readings falls within this definition.

6. Please do not use any cuff other than that supplied by the manufacturer as this may result in 

    inaccurate measurements.

7. For information regarding potential electromagnetic or other interference between the blood 

    pressure monitor and other devices together with advice regarding avoidance of such 

    interference, please see ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is suggested 

    that the blood pressure monitor be kept 10 meters away from other wireless devices, such as 

    WLAN unit, cell phone, microwave oven, etc.

8. This product should not be used as a USB device.

9. If the blood pressure measurement (systolic or diastolic) is outside the rated range specified in 

    part SPECIFICATIONS, the monitor will immediately display a technical alarm on the LED 

    screen. In this case, repeat the measurement ensuring that the proper measurement procedures 

    are followed and/or consult with your medical professional. The technical alarm is preset in the 

    factory and cannot be adjusted or inactivated. This technical alarm is assigned as low priority 

    according to IEC 60601-1-8. The technical alarm does not need to be reset.

10. This device requires a medical AC adapter with an output of DC 5.0V that complies with 

    IEC60601-1/UL 60601-1 and IEC 60601-1-2/EN 60601-1-2 such as ASSA81a-050100

    (input:100-240V, 50/60Hz, 0.45A; output: DC 5V, 1.0A). Please note that the monitor jack size is 

    USB micro-B. The USB jack should be used for charging only.

11. Use of Charging Cable other than those specified or provided by the manufacturer of this 

    equipment could result in increased electromagnetic emissions or decreased electromagnetic 

    immunity of this equipment and result in improper operation.

12. Measurements are not possible in patients with a high frequency of arrhythmias.

13. The device is not intended for use on neonates, children or pregnant women. (Clinical testing 

11

Содержание Neo BP5S

Страница 1: ...Smart Blood Pressure Monitor Tensiom tre connect Sfigmomanometro intelligente Tensi metro inteligente Instruction For Use Notice d utilisation Istruzioni per l uso Instrucciones de uso...

Страница 2: ...EN Smart Blood Pressure Monitor 1 1 1 1 2 2 3 4 5 7 9 10 12 13 15 16 17 18 19 20...

Страница 3: ...itor Electronic Sphygmomanometer is intended for use in a professional setting or at home and is a non invasive blood pressure measurement system It is designed to measure the systolic and diastolic b...

Страница 4: ...your physician for proper interpretation of blood pressure results CONTRAINDICATION It is not recommended for people with serious arrhythmia to use this Smart Blood Pressure Monitor PARTS AND DISPLAY...

Страница 5: ...ith both iOS and Android devices such as iPhone X iPhone 8 Samsung Galaxy S6 Edge SM G9250 Samsung Galaxy Note3 Lite Motorola Nexus 6 For a complete list of compatible devices visit our support on pag...

Страница 6: ...thlabs com and sign in with the same account Power on the Monitor Press START button for at least 3 seconds until the LED screen display all characters to power on the monitor at the first use If cann...

Страница 7: ...lly f Now you can take blood pressure measurement without smartphone and upload data when Bluetooth is connecting MEASUREMENT PROCEDURES Blood pressure can be affected by the position of the cuff and...

Страница 8: ...as a desk or table When the left arm blue arrow points down to the hand is measured position the monitor in the middle of your arm so that it is aligned with your middle finger When the right arm red...

Страница 9: ...commended If the cuff becomes dirty clean it with a moistened cloth Do not rinse the monitor or cuff with running water Left arm measurement Right arm measurement TAKING YOUR BLOOD PRESSURE READING a...

Страница 10: ...ss the START button to turn off the monitor manually Note Please consult a health care professional for interpretation of pressure measurements BLUETOOTH FUNCTION Connect to iOS Android Device via Blu...

Страница 11: ...m 19 mm 5 57 x 2 38 x 0 75 5 Cuff circumference 22cm 42cm 8 66 to 16 54 6 Weight approx 240g 8 47 oz including cuff 7 Measuring method Oscillometric method automatic inflation and measurement 8 Memory...

Страница 12: ...e of a mastectomy d Simultaneous use with other medical monitoring equipment on the same limb e The blood circulation of the user needs to be checked 3 Do not use this product in a moving vehicle as t...

Страница 13: ...echnical alarm on the LED screen In this case repeat the measurement ensuring that the proper measurement procedures are followed and or consult with your medical professional The technical alarm is p...

Страница 14: ...gerous Keep the batteries and the unit out of the reach of children and disabled persons 20 If you are allergic to plastic rubber please don t use this device 21 Attention that changes or modification...

Страница 15: ...zard such as a fire or explosion Do not plug or unplug the power cord into the electrical outlet with wet hands If the AC adapter is abnormal please change the adapter Do not use the monitor while cha...

Страница 16: ...itor by pressing the START button and holding for about 10 seconds then connect the mobile device correctly and try again If the monitor is still abnormal please contact the local distributor or the f...

Страница 17: ...using water diluted disinfectant alcohol or diluted detergent 5 The monitor can maintain the safety and performance characteristics for a minimum of 10 000measurements or three years of usage and the...

Страница 18: ...ur unit WARRANTY INFORMATION iHealth Labs Inc iHealth warrants the iHealth hardware the Product and only the Product against defects in materials and workmanship under normal use for a period of one y...

Страница 19: ...Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your local authority or retailer for recycli...

Страница 20: ...ions not expressly approved by iHealth Lab Inc would void the user s authority to operate the product Note This product has been tested and found to comply with the limits for a Class B digital device...

Страница 21: ...Mark is obligatory for products it applies to the manufacturer affixes the marking in order to be allowed to sell his product in the European market Hereby ANDON HEALTH CO LTD declares that the radio...

Страница 22: ...60 2 2013 Non Invasive Sphygmomanometers Part 2 Clinical Validation Of Automat ed Measurement Type ELECTROMAGNETIC COMPATIBILITY INFORMATION This product is applicable to the equipment and system requ...

Страница 23: ...Table 3 Proximity fields from RF wireless communications equipment 21...

Страница 24: ...Table 4 Input a c power Port 22...