7

d) Simultaneous use with other medical monitoring equipment on the same limb 
e) The blood circulation of the user needs to be checked 
 

3.

 

This Wireless Blood Pressure Wrist Monitor is designed for adults and should never be used 

on infants, young children, pregnant or pre-eclamptic patients. Consult your physician before 
use on children. 

4.

 

Do not use this product in a moving vehicle as this may result in inaccurate measurements. 

5.

 

Blood pressure measurements determined by this product are equivalent to those obtained by 
professional healthcare practitioners using the cuff/stethoscope auscultation method within the 
limits prescribed by the American National Standard, Electronic or Automated 
Sphygmomanometer. 

6.

 

Information regarding potential electromagnetic or other interference between the blood 
pressure monitor and other devices together with advice regarding avoidance of such 
interference, please see ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is 
suggested that the blood pressure monitor be kept 10 meters away from other wireless devices, 
such as WLAN unit, cell phone, microwave oven, etc. 

7.

 

It is recommended that the iOS device be set in Airplane mode during measurement to avoid 
strong magnetism interference. If a call comes in during the measurement, the measurement 
process will be terminated automatically. 

8.

 

If Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol 
will be displayed. Under this condition, the Wireless Blood Pressure Wrist Monitor can keep 
functioning, but the results may be inaccurate. Please consult your physician for accurate 
assessment. 

  There are 2 conditions under which the signal of IHB will be displayed: 

1) The coefficient of variation (CV) of pulse period >25%. 

2) The difference of adjacent pulse period 

0.14s and the number of such pulse takes more than 

53 percent of the total number of pulses.   

9.

 

Please do not use any other cuff other than that supplied by the manufacturer as this may result 
in measurement errors and a biocompatible hazard.   

10.

 

This  product  might  not  meet  its  performance  specifications  if  stored  or  used  outside  the 

specified temperature and humidity ranges. 

11.

 

Please do not share the cuff with any infectious person to avoid cross-infection. 

12.

 

This product should not be used as a USB device. 

13.

 

This product is verified by auscultatory method. It is recommended that you check Annex B of 
ANSI/AAMI SP-10:2002+A1:2003+A2:2006 for verification method details if needed. 

14.

 

If  the  determined  blood  pressure  (systolic  or  diastolic)  is  outside  the  rated  range  specified  in 
part SPECIFICATIONS, the app  will immediately display a technical alarm on screen. In this 
case,  consult  a  physician  or  ensure  that  proper  measurement  procedures  are  followed.  The 
technical  alarm  is  preset  in  the  factory  and  cannot  be  adjusted  or  inactivated.  This  technical 
alarm  is  assigned  as  low  priority  according  to  IEC  60601-1-8.The  technical  alarm  is 
non-latching and does not need to be reset.   

15.

 

A medical AC adapter with an output of DC 5.0V and complies with IEC 60601-1/UL 60601-1 
and  IEC  60601-1-2/EN  60601-1-2  is  suitable  for  this  monitor,  such  as  ASP5-05010002JU 
(input:  100-240V,  50/60Hz,  200mA;  output:  DC 5V,  1.0A).  Please note  that  the  monitor  jack 
size is USB mini B. 
 

 

Содержание BP7

Страница 1: ...OWNER S MANUAL...

Страница 2: ...ARTS AND DISPLAY INDICATORS 3 SET UP REQUIREMENTS 3 SET UP PROCEDURES 4 MEASUREMENT PROCEDURES 4 SPECIFICATIONS 6 GENERAL SAFETY AND PRECAUTIONS 6 BATTERY HANDLING AND USAGE 9 TROUBLESHOOTING 10 CARE...

Страница 3: ...ystolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist The measurement range of the cuff circumference...

Страница 4: ...ne and iPad models iPod touch 4th generation iPod touch 3rd generation iPhone 4S iPhone 4 iPhone 3GS iPad 3rd generation iPad 2 iPad The iOS version of these devices should be V4 0 or higher SET UP PR...

Страница 5: ...DURES Apply the Cuff a Place your hand palm side up in front of you and wrap the cuff around a bare wrist 1 2 cm above the wrist joint b If the monitor is correctly placed the iHealth logo will be fac...

Страница 6: ...lthcare professional for interpretation of blood pressure measurements Measuring without an iOS Device Offline measurement is enabled automatically at your first online measurement To start offline me...

Страница 7: ...quency Band 2 402 2 480 GHz 13 Environmental temperature for operation 5 40 41 104 14 Environmental humidity for operation 90 RH 15 Environmental temperature for storage and transport 20 55 4 131 16 E...

Страница 8: ...but the results may be inaccurate Please consult your physician for accurate assessment There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of variation CV of puls...

Страница 9: ...rsonnel could result in a hazard such as a fire or explosion Do not plug or unplug the power cord into the electrical outlet with wet hands If the AC adapter is abnormal please change the adapter Do n...

Страница 10: ...avoid speaking or movement during the test Bluetooth connection unstable Bluetooth connection unsuccessful BP7 is abnormal or strong electromagnetic interference is present Reset iOS device Reset BP7...

Страница 11: ...will possibly damage your unit 12 It is recommended that if the cuff is used for example in a hospital or a clinic it be disinfected twice a week Wipe the inner side the side that contacts skin of th...

Страница 12: ...Lotus Global Co Ltd 15 Alexandra Road London UK NW8 0DP Tel 0044 20 75868010 Fax 0044 20 79006187 ANDON HEALTH CO LTD No 3 Jinping Street Ya An Road Nankai District Tianjin 300190 China Tel 86 22 6052...

Страница 13: ...le fonctionnement OTHER STANDARDS AND COMPLIANCES The Wireless Blood Pressure Wrist Monitor corresponds to the following standards IEC 60601 1 2006 Medical electrical equipment Part 1 General require...

Страница 14: ...kV for power supply lines 2 kV for power supply lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth...

Страница 15: ...watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a...

Страница 16: ...and mobile RF communications equipment transmitters and the monitor as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter...

Отзывы: