10.6.2 Simulation menu
►
Select [S�FLW] and set the flow value to be simulated�
►
Select [S�TMP] and set the temperature value to be simulated�
►
Select [S�Tim] and set the time of the simulation in minutes�
►
Select [S�On] and set the function:
-On: The simulation starts� The values are simulated for the time set
at [S�Tim]� [SIM] is displayed simultaneously with the process values�
Cancel with [Enter]�
-OFF: The simulation is not active�
Menu SIM:
[S�FLW]
[S�TMP]
[S�Tim]
[S�On]
10.6.3 Reset all parameters to factory setting
►
Select [rES]�
►
Briefly press [Enter]�
► keep [▲] or [▼] pressed.
> [----] is displayed�
►
Briefly press [Enter]�
� We recommend taking down your own settings
in that table before carrying out a reset�
Menu EF:
[rES]
11 Operation
11.1 Reading the process value
The LEDs 1-6 signal which process value is currently displayed in which unit�
The process value to be displayed as standard (temperature, flow velocity or
meter reading of the totaliser) can be preset
�
A standard unit of measurement can be defined for the flow velocity
�
11.2 Changing the process value display in the RUN mode
►
Briefly press
[▼]
or
[▲]
in the RUN mode�
> The unit displays the current measured value in the selected display unit for
approx� 30 s, the corresponding indicator LED lights