![IFM Efector 500 PF2652 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/ifm/efector-500-pf2652/efector-500-pf2652_operating-instructions-manual_3906630020.webp)
• The pressure sensor
detects the system pressure,
• shows the current system pressure on its
display,
• and generates
2 output signals
according to the set output
configuration.
Applications
Type of pressure: relative pressure
Avoid static and dynamic overpressure exceeding the given over-
load pressure.
For gaseous media the application is limited to max. 25 bar.
Even if the bursting pressure is exceeded only for a short time
the unit can be destroyed (danger of injuries)!
21
ENGLISH
Function and features
p-switching (
PnP
)
Output 1
Output 2
Output polarity (applies to both
switching outputs)
Switching function
(output 1 and output 2;
function can be selected for
each output separately)
n-switching (
nPn
)
I
: 4 ... 20 mA
Analogue output
(only output 2)
U
: 0 ... 10 V
hysteresis function / N.O. (
Hno
)
hysteresis function / N.C. (
Hnc
)
window function / N.O. (
Fno
)
window function / N.C. (
Fnc
)
Berstdruck
Permissible
overl. pressure
Measuring range
-1 ... 100
200
650
-0.1 ... 10
20
65
bar
MPa
-10 ... 1 450
2 900
9 425
PSI