7
UK
5 Functions and features
The unit continuously detects the level in tanks.
5.1 Applications
• Water, water-based media
• Oils, oil-based media (only for operation with coaxial probe)
• Compatible with G ¾ process connections
Application examples:
-Detection of cleaning liquid in a parts cleaning system
-Monitoring of hydraulic oil in a hydraulic power unit (only for operation with
coaxial probe)
-Detection of cooling water in an industrial cooling system
-Detection of hot glue in corrugated cardboard manufacture
The unit complies with the standard EN 61000-6-4 and is a class A product.
The unit may cause radio interference in domestic areas. If interference occurs,
the user must take appropriate actions.
The microwave energy radiated by the unit is much below that of mobile
phones.
According to the current state of science the operation of the unit can be
classified to be harmless to human health.
5.2 Restriction of the application area
Incorrect measurements may be caused by the following media:
-Highly absorbing surfaces (e.g. foam).
-Intensely bubbling surfaces.
-Media which are very inhomogeneous, separate from each other thus
forming separation layers (e.g. oil layer on water).
►
Check the function by performing an application test.
►
Installation in a steady area
.
> In case of signal loss, the unit displays [SEnS] and switches the
.
• The unit is not suitable for bulk materials (e.g. plastic granulates).