18
9.4 User settings (optional)
9.4.1 Determine the standard unit of measurement for volumetric flow
► Select [UniF] and set the unit of measurement:
[Lmin], [m3h], [GPm] or [GPh]�
9.4.2 Determine the standard unit of measurement for temperature
► Select [UniT] and set the unit of measurement: [°C] or [°F]�
9.4.3 Setting of the display mode
► Select [diS] and determine the update rate and orientation of the display:
-[d1] = update of the measured values every 500 ms�
-[d2] = update of the measured values every 1000 ms�
-[d3] = update of the measured values every 2000 ms�
-[rd1], [rd2], [rd3] = display as for d1, d2, d3; rotated by 180°�
-[OFF] = the display is switched off in the operating mode�
9.4.4 Customer-specific calibration of the curve of measured values
► Select [CGA] and set the value between 60 % and 140 %
(100 % = factory preset)�
9.4.5 Calibration reset
► Select [CAr]�
► Press [Set] and keep it pressed until [----] is displayed�
► Press [Mode/Enter] briefly�
> Value is reset (CGA = 100)�
9.4.6 Setting the start-up delay
► Select [dSt] and set the numerical value in seconds�
9.4.7 Setting the damping of the measured values
► Select [dAP] and the damping constant in seconds (t value 63 %)�
9.4.8 Setting the error behaviour of OUT1 / OUT2
► Select [FOU1] / [FOU2] and determine the value:
-[On] = output 1 / 2 switches ON in case of an error�
-[OFF] = output 1 / 2 switches OFF in case of an error�