W10MT3A Variac

INPUT POWER: Metered Variac Autotransformers are supplied with a three-wire cord 

and plug for connection to 50-to 60-cycle nominal l20-volt power line. The third wire in 

the cord-plug circuit provides automatic grounding of external metal parts, as a safety 

feature. Do not attempt to operate the units on direct current, or damage will result.

LOADS: Metered Variac Autotransformers are suitable for the control of any alternating-

current load within their rating of 10 amperes   Since they are connected for overvoltage 

(117% of line) operation, care must be taken to avoid damage to loads that are incapable 

of withstanding higher-than-line voltage. Metered models are, of course, particularly 

useful in determining load current, or watts, versus applied voltage.

SWITCHING: To avoid possible destructive surges, set the control knob to the full 

counter clockwise (zero voltage) position before connecting the autotransformer to 

the line, or the load to the autotransformer. Slowly rotate the control knob clockwise 

(increasing the voltage), and observe the meter. Unusually high readings signify trouble; 

the voltage should be returned to zero and the trouble corrected. The meter range is 

changed by a current transformer for  highest possible accuracy; make-before-break 

contacts permit switching under load.

ACCESS: No user-serviceable parts inside.

FUSE: The meter circuit of this unit are fused to prevent overload damage to the meter 

movements. The fuse required is: 12 ampere Bussman Type ABC-12.  These are fast-

blow types; replacement with slow blow fuses or fuses of a higher rating will endanger 

the meters.

iii