Maintenance

4.2.3  Calibration Procedure

To calibrate the 1409, proceed as follows

1.  Determine and employ proper metrological 

practices.

Allow a confi dence band for the uncertainty 
of the measuring instrument and setup.

2.  Set test conditions:

See Calibration conditions in Specifi cations 
Chapter 2 above.
If a Digibridge is used, set to 1 Vrms, series 
model. 
Calibrated value is the capacitance added 
when the standard is plugged into binding 
posts.

3.  Determine the allowable drift limits for the 

capacitance reading.

<0.1% per year

4.  Confi rm that the readings fall within these 

drift  limits,  allowing  for  the  uncertainty 
band.

If the reading falls outside the limits,the unit 
may need to be monitored  or returned to 
IET,

There are no adjustments in the 1409.

4.3  Replaceable Parts List

Reference

IET Pt No

Description

1

32-21

Binding Post, Red

2

32-20

Binding Post, Black

3

32-19

Binding Post, Gold

3

BP-1000-RED

Binding Post, Red

3

BP-1000-BLK

Binding Post, Black

3

BP-1000-GRN

Binding Post, Green

Not Shown

UG-194

bnc connecor

Not Shown

SCA-*-Cap

SCA capacitor assembly

Replace * with nominal capacitance value

4

5

6

Figure 4-1 

10 µF - 1000 µF

 R

eplaceable parts

Содержание 1409 Series

Страница 1: ...SUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC 1409 Series Capacitance Standard Operation Manual Copyright 2022 IET Labs Inc Visit www ietlabs com for m...

Страница 2: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Страница 3: ...at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the speci cations shall void this warranty IET shall...

Страница 4: ...ective 2002 96 EC This product complies with the WEEE Directive 2002 96 EC marking requirements The a xed label indicates that you must not discard this electrical electronic product in domestic house...

Страница 5: ...nstructions contained in the warnings must be followed CAUTION DO NOT Operate in an Explosive Atmosphere Do not operate the instrument in the presence of in ammable gasses or fumes Operation of any el...

Страница 6: ...RE USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when workin...

Страница 7: ...Setup 4 3 2 Connections to Capacitor 4 3 3 Environmental Conditions 4 3 3 1 Operating Temperature 4 3 3 2 Storage Temperature 5 3 4 Shipping and Handling 5 Chapter 4 Maintenance 7 4 1 Preventive Main...

Страница 8: ...vi...

Страница 9: ...t The capacitors are made out of silvered mica and foil spring held in a heavy metal camp for mechanical stability The units are heat cycled for stability placed in a cast aluminum case with silica ge...

Страница 10: ...0 02 20 1 kHz 0 0003 100 ppm 500 10 kHz 100 nF 1409 T 0 02 20 1 kHz 0 0003 100 ppm 500 10 kHz 1 F 1409 Y 0 02 20 1 kHz 0 0002 100 ppm 500 10 kHz 10 F 1409 10 F 0 04 50 100 Hz 0 0005 200 ppm 22 Vrms 1...

Страница 11: ...ered mica and foil pile spring held in a heavy metal clamping structure for mechanical stability Calibration Measured value is capacitance added when standard is plugged directly into binding posts St...

Страница 12: ...F 1409 capacitors have 3 binding posts HI LO and GND as shown in gure 3 1 10 F 1000 F 1409 capacitors have 5 binding posts Sense HI Current HI Sense LO Current LO and GND as shown in gure 3 2 and 3 3...

Страница 13: ...e The 1409 units should be maintained within the storage temperature range of 0 C to 40 C to retain its accuracy within the speci ed limits 3 4 Shipping and Handling The 1409 Series should not be expo...

Страница 14: ...6 1409 Series 6 Operation...

Страница 15: ...pability is available by IET Labs or by a certi ed calibration laboratory If the user should choose to perform this procedure then the considerations below should be observed 4 2 2 General Considerati...

Страница 16: ...owable drift limits for the capacitance reading 0 1 per year 4 Con rm that the readings fall within these drift limits allowing for the uncertainty band If the reading falls outside the limits the uni...

Страница 17: ......

Отзывы: