background image

14

MAINTENANCE

HPRS  Series

Chapter 5

MAINTENANCE

5.1 Preventive Maintenance

The 

HPRS

 Decade Substituter is packaged in a ven-

tilated case.  If it is maintained in a generally clean 
or air-conditioned environment, cleaning will seldom  
be necessary.  In a contaminated atmosphere, cleaning 
may be required.

To maintain optimal accuracy and stability, it is best 
not to open the case of the unit.  In normal service, 
the switches require no additional lubrication.  Dur-
ing the manufacturing process, a light lubrication is 
applied which in most instances is suffi cient for the 
service life of the switches, and yet will not tend to 
collect dust.

5.2 Verifi cation of Performance

5.2.1 Calibration Interval

The HPRS Series instruments should be verifi ed for 
performance at a calibration interval of twelve (12) 
months.  This procedure may be carried out by the 
user, if a calibration capability, is available, by IET 
Labs, or by a certifi ed calibration laboratory.  If the 
user should choose to perform this procedure, then 
the considerations below should be observed.

5.2.2 General Considerations

It is important, whenever testing the 

HPRS

 Series 

Decade Units, to be very aware of the capabilities 
and limitations of the test instruments used.  There 
are a some bridges and direct reading resistance me-

ters or digital multimeters available that can verify 
the accuracy of these units, 

especially

 when used in 

conjunction with standards that can serve to confi rm 
or improve the accuracy of the testing instrument.

Such instruments would have to be 

signifi cantly 

more 

accurate than the specifi ed accuracies for all appli-
cable ranges, in order to perform this task, allowing 
for a band of uncertainty of the instrument itself.  A 
few commercial models, bridges and meters, do exist 
that can do this; consult IET Labs for information.

It is important to allow both the testing instrument 
and the HPRS Substituter to stabilize for a number of 
hours at the nominal operating temperature of 23

0

C, 

and at < 50% RH.  There should be no temperature 
gradients across the unit under test.

Proper metrology practices should be followed in 
performing this verifi cation.

5.2.3 Procedure

1. Confi rm the zero resistance of the unit.

2.  Determine the allowable upper and lower 

limits for each resistance setting of each de-
cade following the specifi ed accuracy given 
in the Specifi cations Section of Chapter 2.

3. Confi rm that the resistances fall within these 

limits.

4. 

If any resistances fall outside these limits, the 
associated switch assembly may be trimmed, 
repaired or replaced.

Содержание HPRS-F-3-0.01

Страница 1: ...TEST AND MEASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC Copyright 2018 IET Labs Inc HPRS im Jan 2018 HPRS SERIES High Power Decade Resistance Substituter User and Service Manual ...

Страница 2: ...Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC ...

Страница 3: ...eplaced at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warranty IET shall not be liable for any indirect special or consequential damages even if notice has been given to the possibility of such damages THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR ...

Страница 4: ...4 HRRS Series Revision History Date Description By March 2013 Initial release LB Jan 2 Major update for standardization of marking and required information in the manual BB ...

Страница 5: ... Category I or it is not marked with a measurement category its measurement terminals must not be connected to line voltage mains and it is not designed for use as for measurements within measurement categories II III and IV The HPRS Series are an indoor use product DANGEROUS PROCEDURE WARNINGS Comply with all WARNINGS Procedures throughout in this manual prevent you from potential hazard These in...

Страница 6: ...nent replacement and internal adjustments must be made by qualified maintenance personnel only DO NOT substitute parts or modify the instrument There are no user serviceable parts inside Refer servicing to qualified personnel Only replaced fuses with correct size and rating to prevent electrical shock To avoid the danger of introducing additional hazards do not install substitute parts or perform ...

Страница 7: ... 45 V ARE USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working with bare conductors or bus bars WHEN WORKING WITH HIGH VOLTAGES POST WARNING SIGNS AND KEEP UNREQUIRED PERSONNEL SAFELY AWAY CAUTION DO NOT APPLY ANY VOLTAGES OR CURRENTS TO THE TERMINALS OF...

Страница 8: ...ctrical Considerations 13 4 4 Dial Setting 13 4 5 Environmental Conditions 13 Chapter 5 MAINTENANCE 14 5 1 Preventive Maintenance 14 5 2Verification of Performance 14 5 2 1 Calibration Interval 14 5 2 2 General Considerations 14 5 2 3 Procedure 14 5 3 Schematic and Replacement Parts 15 5 4 Troubleshooting 15 5 5 Disassembly Component Replacement and Reassembly 15 5 5 1 Disassembly 15 5 5 2 Compone...

Страница 9: ...tion of meters and instru ments HPRS instruments are useful development tools wherever precise resistances with high power handling capacity are required It is important to keep the amount of time power is applied the the HPRS within the duty cycles below The duty cycles assume an ambient temperature of around 230 C Power Time on Time off 101 W continuous none 101 W 5 minutes 10 minutes Exceeding ...

Страница 10: ...01 HPRS F 7 0 1 999 999 9 k 7 0 1 HPRS F 7 1 9 999 999 M 7 1 HPRS F 0 001 99 999 999 k 8 0 001 HPRS F 8 0 01 999 999 99 k 8 0 01 HPRS F 8 0 1 HPRS 200W 9 999 999 9 M 8 0 1 For 0 5 accuracy substitute C for F in part number Options RM for rack mountable version For programmable versions see PRS 300 and PRS 200 202 Accuracy after subtraction of zero resistance at 23 C traceable to SI Option F 1 20 m...

Страница 11: ...zero resistance Zero Resistance 5 mΩ per decade Terminals Two five way binding posts and one ground post electrically connected to case HPRS SERIES HIGH POWER DECADE RESISTANCE SUBSTITUTER WARNING Observe all safety rules when working with high voltages or line voltages Connect the G terminal to earth ground in order to maintain the case at a safe voltage Whenever hazardous voltages 45 V are used ...

Страница 12: ...hipping of the instrument should become necessary 3 2 Installation For a rack mounted model installation in a 19 inch rack may be made using the slots in the rack mounting ears A mounting location that does not expose the unit to excessive heat is recommended For bench models no installation as such is required because this instrument series is not powered Since it is a high accuracy instrument be...

Страница 13: ...0 C Power Time on Time off 101 W continuous none 101 W 5 minutes 10 minutes Exceeding the time on can cause damage to the resis tors If continuous use is required to full rated power see DRS 900 High Power Resistance Substituter 4 3 Electrical Considerations As a good safety practice the case should be grounded using the G terminal Since high voltages may be present it is important to observe all ...

Страница 14: ...apabilities and limitations of the test instruments used There are a some bridges and direct reading resistance me ters or digital multimeters available that can verify the accuracy of these units especially when used in conjunction with standards that can serve to confirm or improve the accuracy of the testing instrument Such instruments would have to be significantly more accurate than the speci...

Страница 15: ...move a decade switch assembly proceed as follows 1 Label and unsolder the bus wires connecting the switch assembly to the resistive loads 2 Loosen the screw holding the knob on the switch shaft and remove the knob 3 Remove the nut holding the switch assembly on the front panel and remove the assembly 4 Replace the assembly by reversing the above steps making certain that the knob screw is aligned ...

Страница 16: ...arranty will be repaired at no charge For any questions on repair costs or shipment instruc tions please contact our Service Department at the above number or visit www ietlabs com to request an RMA online To safeguard an instrument during shipment please use packaging that is adequate to protect it from damage i e equivalent to the original packaging and mark the box Delicate Electronic Instrumen...

Страница 17: ...17 MAINTENANCE HPRS Series Figure 5 1 HPRS Series Schematic Diagram ...

Отзывы: