POCi-W22C-ULT5 Medical Panel PC
Page 20
3.1 Safety Precautions
Please ensure the following safety precautions are adhered to at all times.
External equipment intended for connection to signal input /output or
other connectors, shall comply with relevant UL /IEC standard
(e.g.
IEC60950 -1/IEC62368 -1 for IT equipment and ANSI/AAMI ES60601-1: 2012
AND CAN/CSA- C22.2 No. 60601-1:08/IEC 60601 series for systems–shall
comply with the standard IEC 60601-1-1, Safety requirements for medical
electrical systems. Equipment not complying with UL 60601-1 shall be kept
outside the patient environment, as defined in the standard.
Remove the Power cord form A.C. MAINS if it will not to be used for a
long time.
To prevent the risk of electric shock, make sure power cord is
unplugged from wall socket.
To fully disengage the power to the unit,
please disconnect the power cord from the ac outlet. Refer servicing to
qualified service personnel. The AC outlet shall be readily available and
accessible.
Users must not allow SIP/SOPs and the patient to come into contact at
the same time.
Grounding reliability
can only be achieved when the equipment is
connected to an equivalent receptacle marked “Hospital Only” or “Hospital
Grade”.
Follow the electrostatic precautions
outlined below whenever the
POCi-W22C-ULT5 is opened.
Make sure the power is turned off and the power cord is disconnected
whenever the POCi-W22C-ULT5 is being installed, moved or modified.
Do not apply voltage levels that exceed the specified voltage range
.
Doing so may cause fire and/or an electrical shock. Use a power cord that
matches the voltage of the power outlet, which has been approved and
complies with the safety standard of your particular country.
Electric shocks can occur
if the POCi-W22C-ULT5 chassis is opened when
the POCi-W22C-ULT5 is running. To avoid risk of electric shock, this
equipment must only be connected to a supply mains with protective earth.
Содержание POCi-W22C-ULT5
Страница 11: ...POCi W22C ULT5 Medical Panel PC Page 1 1 Introduction Chapter 1 ...
Страница 25: ...POCi W22C ULT5 Medical Panel PC Page 15 2 Unpacking Chapter 2 ...
Страница 29: ...POCi W22C ULT5 Medical Panel PC Page 19 3 Installation Chapter 3 ...
Страница 43: ...POCi W22C ULT5 Medical Panel PC Page 33 4 BIOS Setup Chapter 4 ...
Страница 77: ...POCi W22C ULT5 Medical Panel PC Page 67 Chapter 5 5 Driver Installation ...
Страница 80: ...POCi W22C ULT5 Medical Panel PC Page 70 Appendix A A Regulatory Compliance ...
Страница 85: ...POCi W22C ULT5 Medical Panel PC Page 75 B Product Disposal Appendix B ...
Страница 87: ...POCi W22C ULT5 Medical Panel PC Page 77 C Maintenance and Cleaning Precautions Appendix C ...
Страница 90: ...POCi W22C ULT5 Medical Panel PC Page 80 Appendix D D Symbol Definitions ...
Страница 92: ...POCi W22C ULT5 Medical Panel PC Page 82 E BIOS Menu Options Appendix E ...
Страница 95: ...POCi W22C ULT5 Medical Panel PC Page 85 Appendix F F Watchdog Timer ...
Страница 98: ...POCi W22C ULT5 Medical Panel PC Page 88 Appendix G G Hazardous Materials Disclosure ...