GeniSPOT Screen Calibration Kit
Page 20
4.3 User Settings
The first time the GeniPASS software is run, all the user settings should be set. After the
initial setup, individual settings can be adjusted if needed. To adjust all the user settings,
for initial setup, follow the instructions below.
Step 1:
Switch to Professional mode to access all the configuration settings
(
Section 5.7
Step 2:
Adjust the user settings indicated below.
S
te
p
0
:
1.
Auto-Dimming Control
configures the auto-dimming controls
(Section 4.4)
2.
Select Monitor
selects the monitor to be tested
(Section 4.5)
3.
Select Performance Criteria
selects the standards for testing the monitor
(Section 4.6)
4.
Profile
sets and adjusts user profile settings
(Section 4.7)
Figure 4-3: User Settings
4.4 Auto-Dimming Control
The
Auto-Dimming Control
window enables, disables and sets the auto-dimming
options.
To adjust the auto-dimming controls, follow these steps:
Step 1:
Click
Auto-Dimming Control
in the main window.
Содержание GeniSPOT
Страница 1: ...GeniSPOT Screen Calibration Kit Page i...
Страница 12: ...GeniSPOT Screen Calibration Kit Page xii List of Tables Table 2 1 Packing List Items 5...
Страница 13: ...GeniSPOT Screen Calibration Kit Page 1 Chapter 1 1 Introduction...
Страница 16: ...GeniSPOT Screen Calibration Kit Page 4 Chapter 2 2 Unpacking...
Страница 18: ...GeniSPOT Screen Calibration Kit Page 6 Chapter 3 3 Installation...
Страница 30: ...GeniSPOT Screen Calibration Kit Page 18 Chapter 4 4 Software Configuration...
Страница 41: ...GeniSPOT Screen Calibration Kit Page 29 Chapter 5 5 Basic Functions...
Страница 56: ...GeniSPOT Screen Calibration Kit Page 44 Chapter 6 6 Advanced Functions...
Страница 61: ...GeniSPOT Screen Calibration Kit Page 49 Figure 6 9 Select Your Monitor...
Страница 62: ...GeniSPOT Screen Calibration Kit Page 50 Appendix A A Compatibility...
Страница 65: ...GeniSPOT Screen Calibration Kit Page 53 Appendix B B Hazardous Materials Disclosure...
Страница 69: ...GeniSPOT Screen Calibration Kit Page 57 Index...
Страница 71: ...GeniSPOT Screen Calibration Kit Page 59...