GeniSPOT Screen Calibration Kit
Page 31
5.2 Conformance Test
The conformance test compares the luminance read by the GeniSPOT to the luminance
that should be produced by the monitor at different luminance levels.
To perform the conformance test, follow these steps:
Step 1:
Click
C
ONFORMANCE
T
EST
in the main window. The conformance test starts
automatically.
Figure 5-2: Conformance Test
Step 2:
Place the GeniSPOT in the white block and click
OK
to continue.
Figure 5-3: Conformance Start
Step 3:
The GeniPASS software displays a white screen, then a black screen and then
progresses from black and through the grays to white. The process takes a few
minutes. No other software should be used while the conformance test is being
performed. The GeniSPOT must remain in the same position at the center of the
monitor throughout the test.
Figure 5-4: Conformance Test Progress
Содержание GeniSPOT
Страница 1: ...GeniSPOT Screen Calibration Kit Page i...
Страница 12: ...GeniSPOT Screen Calibration Kit Page xii List of Tables Table 2 1 Packing List Items 5...
Страница 13: ...GeniSPOT Screen Calibration Kit Page 1 Chapter 1 1 Introduction...
Страница 16: ...GeniSPOT Screen Calibration Kit Page 4 Chapter 2 2 Unpacking...
Страница 18: ...GeniSPOT Screen Calibration Kit Page 6 Chapter 3 3 Installation...
Страница 30: ...GeniSPOT Screen Calibration Kit Page 18 Chapter 4 4 Software Configuration...
Страница 41: ...GeniSPOT Screen Calibration Kit Page 29 Chapter 5 5 Basic Functions...
Страница 56: ...GeniSPOT Screen Calibration Kit Page 44 Chapter 6 6 Advanced Functions...
Страница 61: ...GeniSPOT Screen Calibration Kit Page 49 Figure 6 9 Select Your Monitor...
Страница 62: ...GeniSPOT Screen Calibration Kit Page 50 Appendix A A Compatibility...
Страница 65: ...GeniSPOT Screen Calibration Kit Page 53 Appendix B B Hazardous Materials Disclosure...
Страница 69: ...GeniSPOT Screen Calibration Kit Page 57 Index...
Страница 71: ...GeniSPOT Screen Calibration Kit Page 59...