ZSSC4151Evaluation Kit Start-Up Guide
© 2016 Integrated Device Technology, Inc.
10
March 11, 2016
4. Turn the red knob on the Sensor Replacement Board fully counterclockwise. Select the “P1” field and
click the “Acquire Data” button (see Figure 3.2).
5. Turn the red knob on the Sensor Replacement Board fully clockwise. Select the “P2” field and click the
“Acquire Data” button.
6. Click the Calculate button. The coefficients for the calibration will be calibrated and updated in the
calibration formula textbox at the bottom of the window. HINT: Scroll to the bottom of the textbox so that
the “Output” button appears. Move the cursor over the “Output” button. A popup window will appear
showing the complete set of calibration coefficients.
7. Click the “Write Coefficients To NVM” button. This will download the updated coefficients to the NVM of
the ZSSC4151.
Figure 3.2 Using the Calibration Window Tab to Acquire Data and Calculate Coefficients
7
6
5
4