3-10-1- Result of a calibration
•
Click on
.
•
The list of tests from the personal library is displayed with the status of the last calibration performed for
each reagent lot on-board:
Calibration validated.
Calibration not validated.
Calibration correct with control(s) out-of-range.
•
To display the calibration curve for a test, double-click on the reagent lot number.
•
The software displays the date and the status of the calibration and controls. The value calculated for each
control associated with the calibration (QC1 for Biochemistry; QC1 to QC4 for Immunoassay), is also given.
•
If the calibration is validated, the message
OK
is displayed.
•
In the event of an invalid calibration, a message is displayed at the top of the screen indicating the cause
(see Section 4-1, page 105).
•
When it is not possible to automatically validate a calibration, the
button is displayed when
the user has a Supervisor level of access (or above).
!
When a calibration is displayed before each associated control has been calculated,
and
buttons are not displayed.