4-4- Warning Messages
Warning messages are displayed in the WARNING window, either before the start, or during the run cycle.
If messages a
ff
ecting the operation of the current workload appear, the button is displayed in orange with the
number of messages indicated.
The appearance of certain types of message is also indicated with a flashing icon:
•
A request for calibration, QC1, QC2 or QC3 has been generated automatically.
These requests are generated in the following cases:
•
No calibration in the memory for the analytes requested.
•
Reagent lot present in the compartment is out of date.
•
For Biochemistry if certain Westgard rules have been violated, a request for calibration or control is
generated depending on the rule in question.
•
The calibration and/or control(s) will be carried out after acceptance by the user (by clicking
): If
the user does not want to accept these actions, it is possible to cancel the request by clicking
.
This action will be detailed in the Records.
•
Daily, weekly or monthly maintenance has not been carried out.
Messages are displayed by clicking on the WARNING button. If items are missing, the installation of
the item automatically deletes the message.
!
Depending on the nature of the message, certain tests may stop. In this case, the analyser will
not be able to complete the entire workload. To carry out the entire workload without a
ff
ecting
the deadline for producing the results, it is important to act upon the messages as soon as
possible.
•
The temperatures of the carousel and the reagent compartment are continually monitored. If, during
operations, one of the temperatures is outside the acceptable limits, the analyser will continue to carry
out assays and the icon will be displayed in the SYSTEM STATUS button.
Messages
4-4-
Warning Messages
IDS-iSYS User Manual - Revision M1
Messages 4-4
112
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...