Operator Manual
Aspect 100RTL
L100-3062 REV A
19
OPERATOR CONTROLS
HAND CONTROL
To move tabletop, press appropriate button and hold button depressed until desired position is
reached.
MOVES TOP DOWN
POSITION
TRENDELENBURG
ROLLS TOP RIGHT
PROGRAMMABLE
POSITION 2
SLIDES TOP
FORWARD
1, 2 AND LEVEL
SETUP POSITIONS
SLIDE BACK
BACKWARD
SETS TRENDELENBURG
AND ROLL AT LEVEL
AND ELEVATION DOWN
POSITION 1
PROGRAMMABLE
TRENDELENBURG
SLIDES TOP
ROLLS TOP LEFT
POSITION
REVERSE
MOVES TOP UP
LEVEL
1
idi
LEVEL
S
S
2
ROLL LEFT
1
SLIDE BACK
ROLL LEFT
2
SLIDE FWD
ROLL RIGHT
REV. TREN
HEIGHT UP
TREN
HEIGHT DOWN
REV. TREN
HEIGHT UP
ROLL RIGHT
SLIDE FWD
HEIGHT DOWN
TREN
Programming buttons 1, 2 and LEVEL
(Read through once before attempting).
1)
Use the hand control to place the tabletop in the desired position to be set as a programmed
position (1, 2, or LEVEL). Make sure each axis is correct (trendelenburg, roll, and height).
2)
Wait at least 5 seconds and then press button “S” for about 5 seconds and release.
3)
Quickly, (within about 2 seconds) press the desired setup button (1, 2, or LEVEL). If
programming has been accepted, an audible tone will be heard.
4)
Programming for desired setup is now complete.
5)
Before doing any additional programming or reprogramming, push any of the eight buttons at the
top of the hand control repositioning the tabletop.
6)
Next, push the setup button just programmed.
7)
Go back to step 1 to program or reprogram a button. Before reprogramming a button, the button
must be pushed and the table moved to the set position before making a change to that button’s
programming.
If, over time, an actuator is not returning to its programmed position or an actuator or controller has
been replaced, perform the following steps:
CAUTION
Always check that the patient is secure before
activating any table motion