IMAGING END 

OF TABLETOP 

Aspect 100-4T Q-CARD 

SETUP/PATIENT PREPARATION: 

Moving table (NEVER WITH PATIENT ON BOARD): 

1)  Unlock Rear Casters using Caster Lock Pedal. 
2)  Unlock Front Steering using Front Steer Lock Pedal. 
3)  Position table to desired location. 
4)  Secure table mobility by locking Casters and front steering 

using Rear Caster Lock Pedal and Front Steer Lock Pedal. 

AC Power: 

5)  Plug power cord into properly grounded outlet. 
6)  Protect power cord from damage by casters or other 

equipment. 

Controls: 

1)  Emergency Stop Pushbutton disables all tabletop motion 

control when engaged. 

2)  If tabletop motion controls fail to operate, make sure 

that the Emergency Stop Pushbutton is NOT pushed in. 
 

BATTERY OPERATION (DC POWER): 

1)  The table is FULLY functional, but motions are slower. 
2)  Connect to AC power for a minimum of 8 hours daily to 

 ensure a proper battery charge. 

3)  Failure to push in Emergency Stop Pushbutton when table 

is not in use and not connected to AC power for longer 
than 72 hours, may result in permanent damage to the 
batteries. 
 

CPR: 

Return table to level, retract tabletop to minimize 
overhang and lower table position to a comfortable 
height to perform CPR. 

 

  CLEANING AND STORAGE: 

1)  Unplug power cord and wrap it around 

the cord holder. 

2)  Push in Emergency Stop Pushbutton. 
3)  Lock rear casters and front steering. 
4)  Clean and disinfect table. 

WARNING! Read operator 

manual prior to use 

CONNECTION PORT FOR HAND 
OR FOOT CONTROLS CONNECT 
HERE 

EMERGENCY 
STOP 
PUSHBUTTON 

REAR CASTER LOCK PEDAL 

BRAKES: 

The 100-4T table braking system utilizes a passive magnetic brake; 
it remains engaged in the event of a complete loss of electrical 
power. Power is necessary to release brakes.

  

Braking force is limited by the magnet strength. Use caution when 
applying pressure to the side of the table as the brakes may be 
overcome if more than 50 pounds of force is applied. 

To reduce potential motion during patient transfers with larger 
patients, release the table brake and move the top in the direction 
the patient will be pulled during loading or unloading until the 
tabletop transverse travel limit is reached. Release the panning 
handle to re-engage the lock and then transfer the patient. 

POWER CORD 
HOLDER 

PEDESTAL END OF TABLETOP 

Scan to 
Access 
User 
Manuals 
Online 

FRONT STEER 
LOCK PEDAL 