Operator Manual
Aspect 100-4T G3 and ISR
G3
L100-2843 28-MAR-2018
9
•
Patients must be loaded from the side of table. There is a possibility of tilting,
instability and/or tipping if patient is loaded onto the pedestal end of Table or the
imaging end of the Table.
•
To reduce potential motion during patient transfers with larger patients, move the Table Top
lateral (side to side) in the direction the patient will be moving until the limit of the table top
travel is reached, release the Panhandle Tabletop Float Break to re-engage the break and then
transfer the patient onto the Table.
7.4.
Preparation for Performing CPR
•
Return table to level position, retract tabletop to
minimize overhang and lower table position to a
comfortable height before performing CPR on a patient.
8.
INSTRUCTIONS FOR TABLE OPERATION
8.1.
CASTER MOTION AND STEERING SETTINGS
The motion of the table can be configured three ways so that the rear casters are locked to roll
only forwards or backwards only along the length of the table (“STEER”), be free rolling in all
directions (“FLOAT”) or set so that it will not be able to roll at all (“LOCK”).
WARNING!
Before loading a patient onto the table, always set the
caster motion pedals in the “LOCK” position. Failure to
do so could result in death or serious injury.