KVAR KVA Button
• The 3
φ
KVAR KVA button toggles between the KVAR/KVA
measurements and the PF/kW measurements in the 3
φ
3W
or 3
φ
4W functions.
• Toggling between these measurements can only be done
after the power clamp has calculated the power factor and
kW measurement.
Read Next Button
In 3
φ
3W Function
• Pressing the read next button stores the measured values
for kW
RS(L1L2)
/kVA
RS(L1L2)
and moves to the next measurement.
• Pressing read next after the kW
TS(L2L3)
/kVAR
TS(L2L3)
stores the
values and calculates the kW and PF values for the 3
φ
3W
system.
• After these values have been calculated, press the 3
φ
KVAR
KVA button to display the KVAR and KVA values for the sys-
tem.
• To start another calculation, press the read next button again
to return and take a measurement for the kW
RS(L1L2)
/kVARS
(L1L2)
values.
In 3
φ
4W Function
• Pressing the read next button stores the measured values
for W
R(L1)
/PF
R(L1)
and moves to the next measurement.
• Pressing read next after the W
S(L2)
/PF
S(L2)
stores the values
and moves on to the next measurement.
• Pressing read next after the W
T(L3)
/PF
T(L3)
stores the values
and calculates the kW and PF values for the 3
φ
4W system.
• After these values have been calculated, press the 3
φ
KVAR
KVA button to display the KVAR and KVA values for the sys-
tem.
• To start another calculation, press the read next button again
to return and take a measurement for the W
R(L1)
/PF
R(L1)
values.
6
In any other function
• Pressing the read next button in any other function displays
the measurement values stored with the REC button.
• The REC NO. symbol will be shown on the LCD display and
the record number will flash on the secondary display to noti-
fy the user of what values are being displayed.
• Pressing this button after the meter displays last stored data
point returns to the first stored data point.
• To exit the read next mode, turn the function switch to any
other function. This will exit without losing the stored data.
Kilowatt/Power Factor Measurements
Kilowatt Range
Resolution Accuracy
Voltage/Current
(AC+DC True Power)
Range
0-99.99kW 0.01kW
±(2.0%
+0.05) 600VAC,
800VDC
100-999.9kW
0.1kW
2000A AC/DC
1000-1600kW 1kW
±(2.0%
+5)
Power Factor 0.2-1.0 for 1
φ
2W, 1
φ
3W, 3
φ
3W, and 3
φ
4W
Autoranging Table
0-200VAC
200-600VAC
0-200VDC
200-800VDC
0-200A
0.00-40.00kW 0.00-99.99kW
100.0-160.0kW
200-2000A
0.0-400.0kW 0.0-999.9kW
1000-1600kW
3ØPF
AC+DC
2000KW
2000A
OFF
PFKW
HzV
3Ø3W
3Ø4W
KVARKVA
VA
DC
A/W
ZERO
REC
3
Ø
KVAR
KVA
READ
NEXT
HOLD
CAT III - 600V
Power Clamp
True RMS
61-802