Page 20
Table 14. Temperature Test
Input Reading
Step Temperature
1
-100.0°C
-106.1 to -93.9
2 0.0°C
-3.0
to
3.0
3 100.0°C
96.9
to
103.1
4 1000°C
996
to
1001
Calibration
Calibration of the meter is recommended once a year to ensure performance according to the published
specifications.
Calibration is performed with the use of
CalWin 109VPTCM
Software as illustrated above.
(This software can be obtained by contacting Customer Service at IDEAL INDUSTRIES, INC.)
Do not perform this step if software is already loaded.
The software consists of three files titled
Setup.exe, Data001, and Data002
. After you execute the
setup.exe
program in the Window (95 or 98) environment, you will have
109CalVPTCM100
option in the
PROGRAM set.
Execute
CalWin109VPTCM
choice and you should get the screen shown in Figure 3.
Description of
CalWin109VPTCM
window:
1.
Main Function Indicator:
Displays the function of the Meter.
2.
Main Display:
Displays the main reading of the Meter.
3
. Connection Indicator:
“Connection” shows that the meter is connected to your
computer and the meter and computer are communicating
“Disconnection” shows that the meter and computer are not
connected or are not communicating.
Form number TM61494-5
Rev 2 February 2002