Page 7
Peak Hold:
Peak hold table for Model 61-482 only
Function Range
Accuracy
400mV
Unspecified
4V
±
(1.5% + 300 digits)
40V
400V
DCV
1000V
±
(1.5% + 60 digits)
400mV
Unspecified
4V
±
(1.5% + 300 digits)
40V
400V
ACV
1000V
±
(1.5% + 60 digits)
Function Range
Accuracy
40mA
400mA
±
(3.0% + 60 digits)
DCA
10 A
±
(1.5% + 60 digits)
40mA
400mA
±
(3.0% + 60 digits)
ACA
10 A
±
(1.5% + 60 digits)
Resistance (Ohms) for Model 61-480
Range Resolution
Accuracy
Over Voltage
Protection
300
Ω
.1
Ω
±
(1.0% + 4 digits)
3K
Ω
1
Ω
30K
Ω
10
Ω
±
(0.4% + 3 digit)
600V rms
300K
Ω
.1K
Ω
3M
Ω
1K
Ω
±
(0.6% + 3 digit)
30M
Ω
10K
Ω
±
(1.5% +5 digit)
Open Circuit Voltage:
1.5V approx.
Resistance (Ohms) for Model 61-481-and 61-482
Range Resolution
Accuracy
Over Voltage
Protection
400
Ω
.1
Ω
±
(0.7% + 3 digits)
4K
Ω
1
Ω
40K
Ω
10
Ω
±
(0.4% + 3 digit)
600V rms
400K
Ω
.1K
Ω
4M
Ω
1K
Ω
±
(0.6% + 3 digit)
40M
Ω
10K
Ω
±
(1.5% +5 digit)
Open Circuit Voltage:
1.3V approx.
Form number TM61480-1-2
Rev 5 July 2002