SAFETY 

 
Review the following safety precautions to avoid injury and prevent damage to this product or any 
products connected to it. To avoid potential hazards, use the product only as specified. 
 

CAUTION 

These statements identify conditions or practices that could result in damage to the equipment or other 
property. 
 

WARNING 

These statements identify conditions or practices that could result in personal injury or loss of life. 
 

Specific precautions 
 
Use proper Fuse.

 To avoid fire hazard, use only the fuse type and rating specified for this product. 

 
Do not operate without covers.

 To avoid personal injury, do not apply any voltage or current to the 

product without covers in place. 

 
Electric overload.

 Never apply a voltage to a connector on the product that is outside the range 

specified for that connector. 

 
Avoid electric shock.

 To avoid injury or loss of life, do not connect or disconnect probes or test leads 

while they are connected to a voltage source. 
 

Do not operate in wet/damp conditions.

 To avoid electric shock, do not operate this product in wet or 

damp conditions. 

 

Form number TM61354 

Rev 2 July 2006 

Содержание 61-354 (TRMS)

Страница 1: ...following information Precautions and safety information Specifications Basic maintenance cleaning replacing the battery and fuses Performance test procedures Calibration and calibration adjustment p...

Страница 2: ...7 Certification and compliance 7 Required Equipment 8 Basic Maintenance 9 Opening the Meter Case 9 Replacing the Battery 9 Testing Fuses 10 Fuses Replacement 10 Replacing Fuses 10 Cleaning 10 Perform...

Страница 3: ...ustries com Precautions and Safety Information Use the Meter only as described in the Service Manual If you do not do so the protection provided by the Meter may be impaired Read the Safety Informatio...

Страница 4: ...se proper Fuse To avoid fire hazard use only the fuse type and rating specified for this product Do not operate without covers To avoid personal injury do not apply any voltage or current to the produ...

Страница 5: ...w voltage indicator is indicated Automatic power off time Automatic backslit off 10 minutes Power source 1 5V 2 batteries Maximum input voltage 600V CAT III between V and COM Maximum floating voltage...

Страница 6: ...equency Response 50Hz 500Hz AC Conversion Type AC conversions are ac coupled true rms responding calibrated to the rms value sine wave input Crest Factor C F Peak RMS 1 5 addition error for C F from 1...

Страница 7: ...reading 5 digits 1 5mA 2V For 0 4V 0 8V Overload Protection 600V rms max Continuity Built in buzzer sounds when measured resistance is less than 270 and sound off when measured resistance is more than...

Страница 8: ...of the meter is saved the meter can be turned back on by pressing any switch or changing the rotary switch If the meter is Re Powered by pressing a switch the LCD display the saved state press the Ho...

Страница 9: ...tegory II 1000 Volts CAT III Distribution level mains fixed installation CAT II Local level mains appliances portable equipment Over voltage category CAT I Signal level special equipment or parts of e...

Страница 10: ...mended Model Calibrator AC Voltage Range 0 750V AC Accuracy 0 07 Basic Frequency Range 40 1KHz Accuracy 2 DC Voltage Range 0 1000V DC Accuracy 0 006 Basic Current Range 0 10A Accuracy AC 40Hz to 1KHz...

Страница 11: ...exposed To open the Meter case refer Figure 1 and do the following 1 Disconnect test leads from any live source turn the rotary switch to OFF and remove the test leads from front terminals 2 Remove t...

Страница 12: ...se only a fuse with the amperage interrupt voltage and speed rating specified Fuse rating 10A 500V Fast Replacing Fuses Warning To avoid electrical shock remove the test leads and any input signals be...

Страница 13: ...not necessary to open the case no Adjustments are necessary merely make the required connections apply the designated inputs determine if the reading on the Meter display falls within the acceptable r...

Страница 14: ...0V 500Hz 734V to 766V 6 Turn the rotary switch to V position 7 Set the calibration for the voltage from step 1 to 6 in Table 2 8 Compare the reading on the Meter display with the display reading shown...

Страница 15: ...mpare the reading on the Meter display to the reading in Table 4 Note The meter selects the proper range automatically Each measurement takes about one second per range 2mF takes about 4 5 seconds 4 I...

Страница 16: ...ange shown in Table 6 the meter does not meet specification Table 6 DC Current Test Step Source Reading 1 1 500A 1 483A to 1 517A 2 10 00A 9 88A to 10 12A Testing the Frequency Function To verify the...

Страница 17: ...uit board 10 Using a small flat tipped screwdriver adjust potentiometer VR4 until the display reads 0 000 to 0 003 11 Set the output of AC calibration for 150 0V 100Hz and connect to V Hz and COM inpu...

Страница 18: ...Page 16 VR1 VR67 VR2 VR68 VR66 Figure 5 61 354 Calibration Adjustment Points Form number TM61354 Rev 2 July 2006...

Страница 19: ...Page 17 Form number TM61354 Rev 2 July 2006...

Отзывы: