PERFORMANCE VERIFICATIONS

   

Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning 
correctly.    If the meter does not conform to any of the listed limits, the calibration procedure must be 
performed. 

 

Check the fuse as a BBS, 1A/600V, 10.4*35 (0.41x1.38 inch) UL/CSA, interrupting rating 10kA. 
Check case, leads, display, etc, for any defects. 

 
Table 1 Performance Verification:    Model 61-795   

Function Setting 
/Range 

Apply 

Reading 

Specification 

DCV 

550V DC 

547.2 to 552.8 

[±0.5% ±1 digits] 

ACV 

550V AC 

545.2 to 554.8 

[±0.8% ±4 digits] 

Continuity Ω/

 

<30Ω 

Buzzer sounds 

Not specified 

Resistance: 

 

 

 

LOΩ 

Simulated leads 

.08 to .12 

[±0.02Ω + 0.05 per lead] 

LOΩ 

0Ω 

-0.03 to +0.03 

[±0.03Ω] 

LOΩ 

5Ω 

4.87 to 5.13 

[±2.0% ±3 digits] 

LOΩ 

19Ω 

18.59 to 19.41 

[±2.0% ±3 digits] 

Ω/

 

190Ω 

185.9 to 194.1 

[±2.0% ±3 digits] 

Insulation Resistance: 

 

 

MΩ, 250V 

10MΩ 

250 to 275Vdc 

Source Check 

MΩ, 500V 

10MΩ 

500 to 550Vdc 

Source Check 

MΩ 1000V 

10MΩ 

1000 to 1100Vdc 

Source Check 

MΩ, 250V 
 

0Ω 

0.001 to 0.010 

±10 digits 

100kΩ 

0.090 to 0.110 

±10 digits 

1MΩ 

0.975 to 1.025 

[±2.0% ±5 digits] 

10MΩ 

9.75 to 10.25 

[±2.0% ±5 digits] 

100MΩ 

94.5 to 105.5 

[±5.0% ±5 digits] 

MΩ, 500V 

100kΩ 

0.085 to 0.115 

±15 digits 

1MΩ 

0.975 to 1.025 

[±2.0% ±5 digits] 

10MΩ 

9.75 to 10.25 

[±2.0% ±5 digits] 

100MΩ 

97.5 to 102.5 

[±2.0% ±5 digits] 

1000MΩ 

945 to 1055 

[±5.0% ±5 digits] 

MΩ, 1000V 

100kΩ 

0.085 to 0.015 

±15 digits 

1MΩ 

0.975 to 1.025 

[±2.0% ±5 digits] 

10MΩ 

9.75 to 10.25 

[±2.0% ±5 digits] 

100MΩ 

97.5 to 102.5 

[±2.0% ±5 digits] 

1000MΩ 

965 to 1035 

[±3.0% ±5 digits] 

MΩ, 500V 

500KΩ 

<500V 

Source Check 

Check on each 
Voltage range   

0MΩ 

<3.0mA 

Source Check 

Содержание 61-795

Страница 1: ...61 795 The Service Information provides the following Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the battery and fuses Form number TM61795 Revision 4 Date July 2011 Page i ...

Страница 2: ...erformance Verification 3 Table 1 Performance Verification Model 61 795 3 Calibration 4 5 Table 2 Calibration Range Inputs Model 61 795 5 Replacing the Battery 5 Replacing Fuses 5 Circuit Board Drawing 6 Battery and Fuse Replacement Drawing 7 Revision 4 Remove on page 3 Low ohms source check This check is not needed to verify the performance of the instrument ...

Страница 3: ... are listed in this manual CAUTION These statements identify conditions or practices that could result in damage to the equipment or other property WARNING These statements identify conditions or practices that could result in personal injury or loss of life Specific precautions Use proper Fuse To avoid fire hazard use only the fuse type and rating specified for this product Do not operate without...

Страница 4: ...4 Basic Frequency Range 10 500KHz Accuracy 2 DC Voltage Range 0 1000V dc Accuracy 0 006 Basic Current Range 0 10A Accuracy AC 45Hz to 65Hz 0 05 Basic DC 0 008 Basic Frequency Source 0 01Hz 2 0000MHz Accuracy 0 0025 Amplitude 0 5V p p 1 0V p p square wave Accuracy 5 45Hz 1KHz Ω range 0 01Ω 330M Accuracy 0 007 Basic Capacitance Range 1pF 1 1mF Accuracy 0 19 Basic Temperature Range 200 C 1800 C Accur...

Страница 5: ... 12 0 02Ω 0 05 per lead LOΩ 0Ω 0 03 to 0 03 0 03Ω LOΩ 5Ω 4 87 to 5 13 2 0 3 digits LOΩ 19Ω 18 59 to 19 41 2 0 3 digits Ω 190Ω 185 9 to 194 1 2 0 3 digits Insulation Resistance MΩ 250V 10MΩ 250 to 275Vdc Source Check MΩ 500V 10MΩ 500 to 550Vdc Source Check MΩ 1000V 10MΩ 1000 to 1100Vdc Source Check MΩ 250V 0Ω 0 001 to 0 010 10 digits 100kΩ 0 090 to 0 110 10 digits 1MΩ 0 975 to 1 025 2 0 5 digits 10...

Страница 6: ...oosen the screws from the case bottom Remove the case bottom 5 Short JP1 then replace the case bottom and battery compartment cover Refer to Figure 1 6 Press the key to turn on the instrument The LCD displays the version of the software 7 Push the TEST key to enter the calibration mode 8 Press the key to select the range to calibrate 9 Input the range standard value as listed in Table 2 Push the T...

Страница 7: ... 0MΩ 21 1000V MΩ 10 00MΩ 22 1000V MΩ 1 000MΩ 23 1000V MΩ 0 000MΩ Battery Replacement Refer to Figure 2 1 Disconnect the test leads from any circuit under test 2 Press the key to switch off the instrument 3 Remove the hood and battery compartment cover by using a screwdriver 4 Remove the batteries replacing them with new ones all of the same type 1 5 6 NEDA 15F IEC R6 JIS UM 3 Alkaline batteries ar...

Страница 8: ...er using a screwdriver 6 Check the fuse with a digital multimeter with a source current 10mA in low resistance range Replace fuse with a BBS 1A 600V 10 3 38 1 5x0 41 inch UL CSA interrupting rating 10kA 7 Install input cover and secure with screw 8 Replace the case bottom and secure with screw 9 Replace the battery compartment cover pressing to close it and secure with screw 10 Install the hood ...

Страница 9: ...Form number TM61795 Rev 4 July 2011 Page 7 Figure 1 ...

Страница 10: ...Form number TM61795 Rev 4 July 2011 Page 8 Figure 2 ...

Отзывы: