Page 19
Calibration
Environmental Conditions
Perform calibration at an ambient temperature and relative humidity (23°C±2°C and R.H.
≦
80%). Allow the
instrument to sit at this temperature for at least 30 minutes.
The Operation of Calibration Procedure Mode
Activate Calibration Procedure Mode
Press “
Hold
” button while powering up, then press buttons in the following sequence to activate the
calibration procedure mode
Range → Backlight → VoltSense → Hold → Blue → Min Max
. (
The
“Auto” segment will flash while you are in calibration mode. )
Deactivate Calibration Procedure Mode
Turn the meter off to deactivate calibration procedure mode.
Set Meter Model mode
(61-484 and 61-486 61-486 / 61-484)
After Activate Calibration Procedure Mode, Press the “Min Max” button to select the meter model.
Standard Function Calibration
(Voltage, Ampere, Resistance, Diode, Capacitance)
Dial the rotary switch and press the “
Blue
” button to select the measurement mode.
Press the “
Range
” button to select the range which will be calibrated.
Apply the standard signal indicated in the Calibration Adjustment Point Table to the suitable input
terminals.
Once LCD display to stable, Press the “
Hold
” button to confirm the reading.
If the reading is not withi
n a 10% tolerance of the correct reading when pressing the “Hold” button,
the LCD will display “Err”. Press “Hold” button to exit. Meter will need to be returned for repair.
Temperature Calibration
Connect the temperature calibrator to the meter and wait for at 30 minutes
Apply the standard voltage indicated in the Calibration Adjustment Point Table to the input
terminals.
Once the LCD display is stable, Press the “Hold” button to confirm reading.
If the reading is not within a 10% tolerance of the corr
ect reading when pressing the “Hold” button,
the LCD will display “Err”. Press “Hold” button to exit. Meter will need to be returned for repair.