Page 19
Table 6 Capacitance Test:
Step
Range
Source
Reading
1
40nF
10.000nF
9.60 to 10.40
2
40nF
30.00nF
29.00 to 31.00
3
400nF
300.0nF
290.5 to 309.5
4
4µF
3.000µF
2.905 to 3.095
5
40.00µF
30.00µF
29.05 to 30.95
6
100.0µF
100.0µF
79.5 to 120.5
7
1000µF
1000µF
795 to 1205
Checking the Diode Test Function (for 61-340 / 61-342)
To check the diode test function, do the following:
1. Connect the DVM (Digital Voltage Meter) to the
and
COM
inputs on the meter.
2.
Turn the meter’s rotary switch to
.
Press the
SELECT
button to select the
function
The DVM display should display OL
3. Insert a Si diode with correct polarity “+” and “-“
The meter display should read within 0.5~0.7Vdc
4. Reverse polarity on the Si diode
The meter display should read OL
.
Checking the Continuity Function (for 61-340 / 61-342)
To check the diode test function, do the following:
1. Connect the DVM (Digital Voltage Meter) to the
and
COM
inputs on the meter.
2.
Turn the meter’s rotary switch to
.
a)
Press the
SELECT
button to select the
function
b) The DVM display should display OL
3. Insert a 50
Ω
resistor
a) The meter inner beeper will sound
Form Number
2
TM61340-2
Rev
November 2007