Form Number TM61320-2-4
Rev 4 February 2008
Page 4
Measurement Characteristics
Accuracy is ±(% r number of digits) at 23°C ± 5°C, less than 80% R.H.
(1) DC Volts
Range Resolution
Accuracy
Over Voltage
Protection
400.0mV 100
μ
V
4.000V 1mV
40.00V 10mV
400.0V 100mV
1000V 1V
±(0.5% r 2 digit)
1000V rms
Input Impedance:
10M
Ω
(over 1000M
Ω
in 400mV range).
(2) AC Volts
Range Resolution
Accuracy
Over Voltage
Protection
400.0mV 100
μ
V
Unspecified
4.000V 1mV
±(1.3% r 5 digits)
*
1
40.00V 10mV
400.0V 100mV
750V
1V
±(1.2% r 5 digits)
*
2,3
1000V rms
Input Impedance:
10M
Ω
// less than 100pF.
*
1
Frequency Response:
50Hz
~
300Hz
*
2
Frequency Response:
50Hz
~
500Hz
CMRR / NMRR: (Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
V
AC
: CMRR > 60dB at DC, 50Hz / 60Hz
V
DC
: CMRR > 100dB at DC, 50Hz / 60Hz
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type: 61-320:
Average sensing rms indication calibrated to the sine wave input.
61-322 / 61-324:
AC conversions are ac-coupled, true rms responding,
calibrated to the sine wave input.
*
3
61-322 / 61-324 The specified accuracy is for sine wave at full scale and non-sine wave at half
scale with crest factor up to 2.
Crest Factor:
C.F. = Peak/RMS
+1.5% addition error for C.F. from 1.4 to 3
+3.0% addition error for C.F. from 3 to 4.