33
I N S T A L L A T I O N
E N
5.6
Calibration of Axes
Following the set-up and connection to the PC, the
desktop scanner must be calibrated using the sup-
plied calibration model.
▷
Mount spacer plate No. 1, the fixation plate with
the clamp and the model plate (see
section 4.4.2).
Fig. 5
Calibration model
▷
Insert the calibration model
▷
Create a new patient case and start the scanner
software from the data base.
▷
Cancel the subsequent command prompt for
the 2D scan.
▷
Click on
Scanner Options > Axis calibration to
start the axis calibration.
▷
Follow the software instructions.
If required:
▷
Re-enter the calibration model values already
entered during the installation by clicking on
Application > Register Calibration Object .
5.7
Calibrating the Artex fixator
(accessory)
Before the Artex Fixator is used in the scanner, it
must be calibrated with the Splitex calibration key:
Fig. 6
1
Fixator, bottom element
2
Adhesive surface
3
Splitex plate
4
Splitex calibration key
5
Fixator, upper element
6
Lateral screw
7
Rear screw
▷
Open the Fixator.
▷
Place the Splitex calibration key
4
onto the
upper element
5
.
▷
Place the Splitex plate
3
onto the calibration
key.
▷
Apply adhesive (Loctite) to bottom side of the
Splitex plate.
The scanner should be recalibrated at reg-
ular intervals.
The calibration of the fixator with the Spli-
tex calibration key should be checked at
regular intervals.
1
2
3
4
5
6
7
OBJ_DOKU-0000004530-001.fm Page 33 Tuesday, March 10, 2015 5:32 PM