4.8 PATIENT PENDANT JACK ASSEMBLY REPLACEMENT
Technical Service Manual
4 - 7
LifeCare PCA
4.8
PATIENT
PENDANT
JACK
ASSEMBLY
REPLACEMENT
Instead of the procedure provided in
Section 7.3.9.9, Patient Pendant Jack Assembly
, follow the instructions provided below.
Recommended tools for this procedure are a 1/2 inch nutdriver, diagonal cutters,
and cable ties.
Replacement parts for this procedure are:
Assembly, Patient Pendant Jack
Tie, Cable
To replace the patient pendant jack assembly, see
,
and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosure assemblies as described in
3. Using the nutdriver, remove the two retainer nuts and flat washer that attach the
patient pendant jack to the rear enclosure.
4. Using the diagonal cutters, remove the cable ties that secure the patient pendant
jack assembly to the rear enclosure harness assembly.
5. Remove the spiral wrap and the patient pendant jack assembly.
6. Install the replacement patient pendant jack assembly in the exact reverse order
of removal. Secure the patient pendant jack harness to the rear enclosure harness
assembly with the spiral wrap and cable ties.
7. Reassemble the infuser in the exact reverse order of disassembly.
8. Connect the infuser to AC power.
To verify successful replacement of the patient pendant jack assembly, perform the PVT
in
4.9
CSA
MARK
Instead of the mark shown on the back cover of this manual, the following mark/
description applies:
CAUTION:
Do not bend the patient pendant cord near the patient pendant
or connector jack at an angle of 90
o
or greater.
The ‘C’ and ‘US’ indicators adjacent to the CSA Mark signify that the
product has been evaluated to the applicable CSA and ANSI/UL
Standards, for use in Canada and the U.S., respectively. This ‘US’
indicator includes products eligible to bear the ‘NRTL’ indicator.
NRTL (National Recognized Testing Laboratory), is a designation
granted by the U.S. Occupational Safety and Health Administration
(OSHA) to laboratories which have been recognized to perform
certification to U.S. Standards.
UL 60601-1
CSA 601.1
MCN 160992
US
C
®
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...