ICS Schneider Messtechnik PT130 Скачать руководство пользователя страница 2

From lab to production,  
providing a window into the process 

OPERATING PRINCIPLE 

The PT100 Series Industrial Transducer is used to 
make pressure measurements for hydraulic 
applications and corrosive media up to 10,000 psi. 
These models incorporate a four-arm, 350-ohm bonded 
foil Wheatstone bridge strain gage. An amplifier can be 
added to this proven technology to provide a high-level 
output signal (e.g. 0 – 5 V dc or 0 – 10Vdc) proportional 
to the pressure (within the specified error band).  

SHUNT CALIBRATION 

Internal 

– All 100 Series transducers contain an 

internal Shunt Calibration resistor, R-CAL, which 
provides a simulated output of 80% of full scale when 
pins E and F are shorted.  

External 

– The model PT130 also provides an output of 

80% of full scale when a 34,935-ohm Shunt Calibration 
resistor is placed across pins B and D. External R-CAL 
feature is not available on models PT160 and PT150.  

NOTE: 

The internal and external R-CAL should not be 

used at the same time because incorrect readings will 
result.  

PRESSURE FITTING 

The pressure fitting on all 100 Series transducers 
contains a 1/8”-27 NPT internal pressure port. The 
fitting is fabricated from high-strength stainless steel. 

ELECTRICAL CONNECTORS 

 

All models are equipped with a six pin, Bendix PT02A-
10-6P receptacle. The customer supplied mating plug is
a socket, Bendix PT06A-10-6S, or equivalent.

ELECTRICAL CHARACTERISITCS 

Function 

PT130 

PT150 and PT160 

4 wire 

3 wire 

 

Signal – 

Exci 

Excitation – 

R-Cal Int.

E, F 

E, F 

E, F 

R-Cal Ext.

B, D 

Red 

Signal (+) 

Black 

Signal (-) 

White 

Excitation (+) 

Green 

Excitation (-) 

Blue/Orange 

R-Cal

Model 

Wiring 

Signal 

Excitations 

PT130 

4 wire 

3 mV/V 

10 V dc Recommended, 
15 V dc Max. 

PT150 

4 wire 
3 wire 

0-5 V dc
1-6 V dc

12-32 V dc ± 12 to ± 16 V dc
12-32 V dc

PT160 

4 wire 
3 wire 

0-5 V dc
1-6 V dc

16-32 V dc, ± 14 to ± 16 V dc
16-32 V dc

 

WIRING

 

CODES

 

WIRING COLOR CODE 

TRANSDUCER REPAIR 

Questions concerning warranty, repair cost, delivery, 
and requests for a RA# should be directed to the 
Dynisco Repair Department, 508-541-9400 or email: 
repair @dynisco.com. Please call for a return 
authorization number (RA#) before returning any 
product. Damaged transducers should be returned to: 

Dynisco Instruments  

Attn: RA #_____________  

38 Forge Parkway  

Franklin, MA 02038  

TECHNICAL ASSISTANCE 

 

Please call 800-221-2201 or 508-541-9400 or fax 
508-541-9206.

WARRANTY

This Dynisco product is warranted under terms and 
conditions set forth in the Dynisco Web Pages. Go to 

www.dynisco.com 

and click on “Instruments”. Click 

“Product Warranty” at the bottom of any page for 
complete details.

Отзывы: