ICS Schneider Messtechnik induQ IVMZ.2 Series Скачать руководство пользователя страница 9

 — 

IVMZ

.2 

Installation of 

IVMZ

.2 

Technical changes reserved 

- 9 -

4.1  Installation instructions 

CAUTION! Risk of malfunction due to external mag-

netic fields! 

Magnetic fields close to the device can cause malfunctions and 

should be avoided. 

Ensure that no external magnetic fields are present at the

instal-lation site of the 

IVMZ

.2.

The 

IVMZ

.2 can always be installed anywhere along the

pipeline.

 

However, straight sections of piping are preferable.

Installation can occur in horizontal and vertical pipes. The flow sensor is only suitable for

application in completely filled pipe systems.

As a matter of principle, magnetic inductive flow sensors are widely independent from the

flow profile. An inlet section is not necessary.

To reach a most highly accuracy of the measurement, you should use straight inlet and

outlet sections according to the nominal width (DN). The inlet section has to be at least

10 x DN; the outlet section 5 x DN in order to achieve the specified accuracy.

The inlet and outlet sections and the gaskets must have the

same or a slightly larger inside diameter than the measuring

tube in order to achieve the specified accuracy.

If two or more 

IVMZ

.2 devices are used side by side, maintain

a separation of at least 2.5 cm between adjacent devices.

If adjacent devices are too close together, operation of both

devices may be impaired due to mutual interference.

 

Bubble-formation / run empty possible 

Dirt deposits 

possible 

Run empty 

possible 

Dirt deposits 

possible 

Содержание induQ IVMZ.2 Series

Страница 1: ...Magnetic Inductive Flow Sensor Series IVMZ 2 Types VMZ03 VMZ06 VMZ08 VMZ15 VMZ20 VMZ25 Operating manual Operating manual page 1 20...

Страница 2: ...function 7 4 Installation of IVMZ 2 8 4 1 Installation instructions 9 4 2 Assembly 10 5 Electrical connection 11 5 1 Wirings 12 6 Commissioning and measuring operation 13 6 1 Commissioning 13 6 2 Swi...

Страница 3: ...fects or cause considerable damage to property CAUTION Electric current This sign indicates dangers which could arise from handling of electric current CAUTION Material damage This sign indicates acti...

Страница 4: ...e been carefully checked for their operational reliability before shipment Immediately after receipt please check the outer packaging for damages or any signs of improper handling Report any possible...

Страница 5: ...re that the measuring tube of the IVMZ 2 is always completely filled 4 1 Installation instructions Ensure that the conductivity of the fluid is at least 20 S cm 1 3 Exclusion of liability We accept no...

Страница 6: ...g national regulations for accident prevention and safety in the workplace must be complied with Any internal regulations of the operator must also be complied with even if these are not mentioned in...

Страница 7: ...the middle of the measuring tube between the earthing sleeves The IVMZ 2 does not need any moving parts to make measurements The inside of the meas uring tube is completely open allowing the fluid to...

Страница 8: ...alling check that the wetted materials of the device are suitable for the media being used 9 2 Materials table the equipment is switched off and is in a safe and de energised state the equipment is de...

Страница 9: ...rs are widely independent from the flow profile An inlet section is not necessary To reach a most highly accuracy of the measurement you should use straight inlet and outlet sections according to the...

Страница 10: ...irt deposits Install the appropriate screwed connections at the installation location Insert the IVMZ 2 together with the gaskets Screw the union nuts of the screwed connection onto the pro cess conne...

Страница 11: ...AUTION Electric current The electrical connection should only be carried out by a fully qualified electrician De energize the electrical system before connecting the IVMZ 2 Connection cable Suitable c...

Страница 12: ...ype plate Supply voltage IVMZ 2 with frequency output Push Pull NPN Open Collector PNP Open Collector 1 Push Pull switching outputs of several IVMZ 2 may not be connected in parallel 2 Recommendation...

Страница 13: ...Switch on the supply voltage The green LED lights up The IVMZ 2 is ready and goes into measuring mode 6 3 Measuring operation In the measuring mode the green LED flashes proportional to the measured...

Страница 14: ...vices If necessary the customer must neutralise or purge the unit before return Costs incurred due to inadequate cleaning of the device and possible costs for disposal and or personal injuries will be...

Страница 15: ...ised state the equipment is depressurised and has cooled down Disassembly Remove the electrical connectors Remove the IVMZ 2 using suitable tools Disposal Compliant with the Directives 2011 65 EU RoHS...

Страница 16: ...eristics Frequency output Pulse rate K Factor pulses l 10 000 4000 1000 400 200 80 Resolution ml pulse 0 1 0 25 1 0 2 5 5 0 12 5 Signal shape Square wave signal duty cycle 50 50 Push Pull NPN open col...

Страница 17: ...ure 15 60 C Storage temperature 15 60 C Nominal diameter DN 3 DN 6 DN 8 DN 15 DN 20 DN 25 Inner diameter 3 mm 8 x 2 mm 8 mm 14 mm 18 mm 25 mm Max working pressure at C 10 bar 20 C 8 bar 40 C 6 bar 60...

Страница 18: ...nsions IVMZ 2 DN 3 DN 6 DN 8 DN 15 DN 20 VMZ06 Type L1 L2 D1 D2 VMZ03 85 13 G B 3 VMZ06 85 13 G B 8 x 2 VMZ08 85 13 G B 8 VMZ15 90 16 G B 14 VMZ20 90 16 G1 B 18 IVMZ 2 DN 25 L1 58 M12x1 13 5 D2 89 36...

Страница 19: ...IVMZ 2 Technical changes reserved 19 For your notes...

Страница 20: ......

Отзывы: