– Service Manual
Table 2 continued. . .
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment –
Guidance
Surge IEC
61000-4-5
±
1 kV
differential
mode
±
2 kV
common mode
±
1 kV
differential
mode
±
2 kV
common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage
dips, short
interrup-
tions and
voltage
variations
on power
supply
input lines
IEC
61000-4-11
<
5% U
T
(
>
95% dip in
U
T
) for 0.5
cycle.
40% U
T
(60%
dip in U
T
) for
5 cycles.
70% U
T
(30%
dip in U
T
) for
25 cycles.
<
5% U
T
(
>
95% dip in
U
T
) for 5 sec.
<
5% U
T
(
>
95% dip in
U
T
) for 0.5
cycle.
40% U
T
(60%
dip in U
T
) for
5 cycles.
70% U
T
(30%
dip in U
T
) for
25 cycles.
<
5% U
T
(
>
95% dip in
U
T
) for 5 sec.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the
i
CR 3600 requires
continued operation during
power mains interruptions, it is
recommended that the
i
CR 3600
be powered from an
uninterruptible pwoer supply or a
battery.
Power
frequency
(50/60Hz)
magnetic
field IEC
61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
continued on next page. . .
c
2007-2009
Confidential and Proprietary
Property of
i
CRco, Inc.
vi of 61
Document # 3600-02A Rev C
March 22, 2010