D700 | D720 Series Operating Manual
9
Menu Structure
The following is the menu structure of the instrument:
Item
Display
Selection
Sub Menu
Configure
Sensor
Type
pH
ORP
Unit
pH
mV
Temperature
Temperature Sensor
NTC2.252 kΩ
NTC10 kΩ
Pt100
Pt1000 (Default)
Temperature Offset
0.0000
Temperature Input
Automatic
Manual
Temperature Unit
°C
°F : Default
Calibration
Standard
Solution
Calibration
USA : 7.00,4.01,10.01
[ ] 7.00 [ ] 4.01 [ ] 10.01
NIST : 6.86,4.01,9.18
[ ] 6.86 [ ] 4.01 [ ] 9.18
ORP Standard Solution
235mV (Default)
Field Calibration
Field Calibration
Enter Value
Offset Adjustment
Enter Value
Slope Adjustment
Enter Value
Calibration
Adjustment
Offset Adjustment 1
Offset 2 Adjustment
Slope Adjustment 1
Slope 2 Adjustment
Set Points
Relay 1
Status
ON
OFF
High/Low Setpoint
High Alarm
Low Alarm
Clean
Limit Value
0.00 (Default)
Hysteresis
00.000
Relay 2
Status
ON
OFF
High/Low Setpoint
High Alarm
Low Alarm
Clean
Limit Value
0.00 (Default)
Hysteresis
00.000
Relay 3
Status
ON
OFF
High/Low Setpoint
High Alarm
Low Alarm
Clean
Limit Value
0.00 : Default
Hysteresis
00.000
Содержание ProCon D700 Series
Страница 1: ...D700 D720 Series pH ORP Digital Analyzer Controller Operating Manual ...
Страница 14: ...D700 D720 Series Operating Manual 14 ...
Страница 15: ...D700 D720 Series Operating Manual 15 Measurement Settings ...
Страница 16: ...D700 D720 Series Operating Manual 16 SimplCal Calibration ...
Страница 17: ...D700 D720 Series Operating Manual 17 Field Calibration ...
Страница 18: ...D700 D720 Series Operating Manual 18 Calibration Adjusment ...
Страница 19: ...D700 D720 Series Operating Manual 19 Alarm Relay Settings ...
Страница 20: ...D700 D720 Series Operating Manual 20 Output Settings ...