C700 | C720 Series Operating Manual
11
Factory Default Current Outputs
Cell
Models
Range
Cell 0.01
C250
0 – 2µS
Cell 0.01
C250
0 – 10µS
Cell 0.1
C350
0 – 20µS
Cell 0.01
C450, C650, C950
0 – 100µS
Cell 0.1
C350
0 – 200µS
Cell 0.1
C450, C650, C950
0 – 1,000µS
Cell 0.1
C350
0 – 2,000µS
Cell 1.0
C450, C550, C650, C950
0 – 10,000µS
Cell 10
C450, C650, C950
0 – 200,000µS
Cell 0.4
C554, C850
0 – 500,000µS
Toroidal
C750
0 – 2,000mS
Calibration
Press
[MENU] to enter the calibration menu and select
Calibration
Standard Calibration
Calibration Point 1
Enter given standard liquid value: Defa
Calibration Point 2
Enter given standard liquid value (Example: 84μS/cm)
Calibration Point 3
Enter given standard liquid value (Example: 1413μS/cm)
Calibration Point 4
Enter given standard liquid value (Example: 12.88mS/cm)
Calibration Point 5
Enter given standard liquid value (Example: 111mS/cm)
Field Calibration
Field Calibration
Enter Value
Offset Adjustment
Enter Value
Slope Adjustment
Enter Value
Calibration of Standard Solution
The sensor has been factory calibrated before delivery and users can use it directly.
Note: Calibration if required can be made using at least one of the five calibration points. Not all points are
required for sensor calibration
If calibration is required, prepare a suitable standard reference liquid with known range value. Press the
[MENU] button to enter the
Setting Mode — select the calibration point.
Enter the corresponding reference calibration value, then press
[ENT] key to confirm and enter the calibration screen.
(If the instrument has been calibrated, the screen shows the calibration status. press the
key to switch the calibration state of the
calibration point. If re-calibration is required in this mode, press the
[ENT] key again to enter the re-calibration.)
In the standard solution calibration mode, press the
key to switch the calibration points, press the
[ENT] key to start calibrating.
If prompted to enter the calibration security password, press the
key or
key to set the calibration security password, and
then press the
[ENT] key to confirm the calibration security password.
Voltage:
Point 1
0
µS/cm
Done
-5 mV
Conductivity / TDS / Salinity / Resistivity
Calibrating
Calibration
Sensor Voltage
Successful Calibration