C400 Series Operating Manual
11
Calibration point 1: After entering the calibration mode, the current measurement value will be displayed. This is known as the standard
solution value.
Place the electrode into the standard solution of the corresponding value, and the corresponding voltage mV value and calibration state
will be displayed on the screen.
After completion of calibration,
Done
will be displayed on the right side of the screen.
If you want to calibrate the next point, press the
key to switch the calibration point.
If only one point calibration is required, press the
[MENU] key to return to the superior menu after calibration.
During the calibration process, the calibration status will be displayed on the right side of the screen.
Done
indicates successful calibration,
Calibrating
indicates in calibration, and
Error
indicates a sensor or a calibration issue.
Field Calibration — Select on-site calibration methods:
[Linear calibration]
,
[Offset adjustment]
,
[linear adjustment]
.
Field Calibration
When the data from laboratory or portable instrument are input into this item, the instrument will automatically correct the data.
Conductivity / TDS / Salinity / Resistivity
pH / ORP
µS/cm: 07.00
Vol: -002
Temp: 25.0 °C
Slope: 00.984
Field Calibration
Calibrating
Done
Offset adjustment
Compare the data of laboratory or portable instrument with the data of instrument measurement,if there are errors, the error data can be
modified by this function.
Linear adjustment
Linear values after “field calibration” are stored in this item, with factory data of 1.00. The main value of the instrument displays the known
standard liquid value point.
Place the electrode into the standard solution of the corresponding value, and the corresponding voltage mV value and calibration state
will be displayed on the left side of the screen.
After completion of calibration,
Done
will be displayed on the right side of the screen. If the next point is calibrated, press
to switch the
calibration point.
If only one point calibration is needed, after the calibration is completed, press
[MENU] to exit. During the calibration process, when the
standard solution is wrong, the screen will show an
Error
.
No Conductivity Module Required
Relay
Alarm/Cleaning
Relay 1
Pump/Valve
Relay 2
Pump/Valve
Power
(100-230 VAC)
4-20mA
4 wire
Controller
Conductivity
Sensor
ENT
MENU
ESC
83.01
µS/cm
Conductivity / TDS / Resistiv
ity / Salinity
249
mV
25.0
°C
SP1
SP2
C1
14.51 mA
R
C400