C400 Series Operating Manual

27

◀◀

|

Troubleshooting

1. 

LCD display is not bright

 

Possible Causes:

 Instrument or LCD Screen power supply failure.

 

Solutions:

 Check whether the power supply is connected or not, and check whether the power supply wire of the sensor is connected in 

the wrong direction.

2. 

No current output

 

Possible Causes:

 It could be a fault in the current module or a wiring fault.

 

Solutions:

 Please check that the current output wiring is correct. Please refer to the wiring terminal diagram in the instructions.

3. 

The output current of the transmitter does not match the display current.

 

Possible Causes:

 Current output may not be correctly calibrated.

 

Solutions:

 Please re-calibrate the 20mA output.

4. 

There is a

 

 

on the screen.

 

Possible causes:

 The transmitter does not communicate properly with the sensor.

Solutions:

 Check whether the signal line of the sensor is 

correct. Please refer to the wiring diagram in the instruction.

5. 

Measurements display results fluctuate greatly.

 

Possible causes:

 Sensor wiring error or low display rate setting.

 

Solutions:

 Check wiring or increase display rate appropriately.

Warranty

Icon Process Controls

 warrants this product to be free from significant deviations in material and workmanship for a period of one year 

from the date of purchase. If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return 

to 

Icon Process Controls

 and amendment will be made without any charge. 

We Instruments

 Customer Service Center will determine if 

product problem is due to deviations or customer abuse. Out of warranty products will be repaired on a charge basis.
Authorization must be obtained from 

Icon Process Controls

 Customer Service Center to issue a RIR number before returning items for 

any reason. When applying for authorization, please include date requiring the reason of return. Instruments must be carefully packed to 

prevent damage in shipment and insured against possible damage or loss. 

Icon Process Controls

 will not be responsible for any damage 

resulting from careless or insufficient packing.

Warning:

 Damage as a result of inadequate packaging is the User / distributor’s 

 

responsibility. Please follow the guidelines below before transporting.
Use the original packaging material if possible, when transporting back the unit for repair. Otherwise wrap it with bubble pack and use a 

corrugated box for better protection. Include a brief description of any faults suspected for the convenience of Customer Service Center, if 

possible. If there are any questions, feel free to contact our Customer Service Center or distributors.

by

Phone:

 905.469.9283  •  

Sales:

 [email protected]  •  

Support:

 [email protected]

Содержание ProCon C400 Series

Страница 1: ...C400 Series Controller Conductivity TDS Resistivity Salinity Operating Manual...

Страница 2: ...e 8 Calibration 10 Field Calibration 11 Measurement Settings Quick Start 12 Temperature Calibration 13 Point Calibration 14 Field Calibration 15 Relays 16 4 20mA Ouput 17 RS485 Modbus 18 System Setup...

Страница 3: ...een stored for a long time in an environment where the temperature exceeds 70 C The analyzer must be installed by licensed professionals in accordance with relevant local specifications and instructio...

Страница 4: ...0 01ppt Working Temperature 32 302 F 0 150 C Temperature Compensation Automatic or Manual Output Two 4 20mA 20 4mA 0 20mA Communication RS485 MODBUS RTU Relay Control Contact 5A 240VAC 5A 28VDC Power...

Страница 5: ...on CELL 1 Conductivity Sensor Input 1 NO 1 High set relay working position CELL 2 Conductivity Sensor Input 2 COM 1 Alarm relay common CELL 3 Conductivity Sensor Input 3 NC 1 High set relay resting po...

Страница 6: ...tallation Wall Mounting 53 mm 18 4 mm 5mm Fixture Dimension Fixed bracket M4 screw is installed on the wall first Insert the instrument into the fixture along the chute a b Insert into the fixture alo...

Страница 7: ...nication Warning Conductivity TDS Salinity Resistivity 83 01 S cm 25 0 C C1 14 51 mA SP1 SP2 Measurement Mode Conductivity TDS Salinity Resistivity Calibration Mode V oltage P oint 1 0 S cm Done 5 mV...

Страница 8: ...Temperature Temperature Sensor NTC2 252 k NTC10 k Pt100 Pt1000 Default Temperature Offset 0 0000 Enter Value if Required Temperature Input Automatic Manual Temperature Unit C F Default Calibration St...

Страница 9: ...ault 0 20mA 20 4mA Upper Limit 00 00 Lower Limit 00 00 Current 2 Channel Main Temperature Output Option 4 20mA Default 0 20mA 20 4mA Upper Limit 00 00 Lower Limit 00 00 Output RS485 Baud Rate 4800BPS...

Страница 10: ...he sensor has been factory calibrated before delivery and users can use it directly Note Calibration if required can be made using at least one of the five calibration points Not all points are requir...

Страница 11: ...ect the data Conductivity TDS Salinity Resistivity pH ORP S cm 07 00 Vol 002 T emp 25 0 C Slope 00 984 Field Calibration Calibrating Done Offset adjustment Compare the data of laboratory or portable i...

Страница 12: ...ress ESC key for Back MENU ESC Press ENT key for Save Press ESC key for Back MENU ESC Press ENT key for Save Press ESC key for Back Press ENT key Press ENT key Select Configure Press key Select Sensor...

Страница 13: ...ey for Back MENU ESC Press ENT key for Save Press ESC key for Back MENU ESC Press ENT key for Save Press ESC key for Back Configure MENU ESC Press MENU key Press ENT key Select Configure Press key Pre...

Страница 14: ...e Press ENT key for Save Press ESC key for Back Calibrating Select Value MENU ESC 0000 Enter Value Press ENT key for Save Press ESC key for Back Calibrating Select Value MENU ESC 0000 Enter Value Pres...

Страница 15: ...ess key Press ENT key Select Field Cal Press key Press ENT key Select Offset Press key Press ENT key Select Slope Press key MENU ESC 0000 Enter Value Press ENT key for Save Press ESC key for Back Cali...

Страница 16: ...y Press key Press ENT key Select Status Press key Press ENT key Select High Low Set Press key Press ENT key Select Limit Value Press key Press ENT key Select Hysteresis Press key MENU ESC Press ENT ke...

Страница 17: ...key Press ENT key Select Output Press key Press ENT key Select Current Press key Press ENT key Select Output Opt Press key Press ENT key Select Upper Limit Press key Press ENT key Select Lower Limit...

Страница 18: ...NT key Select Baud Rate Press key Press ENT key Select Parity Check Press key Press ENT key Select Stop Bit Press key Press ENT key Select Network Press key MENU ESC Press ENT key for Save Press ESC k...

Страница 19: ...te Time Press key Press ENT key Select Item Press key Press ENT key Select Item Press key Select Item Press key Press ENT key Select Display Speed Press key Press ENT key Select LED Press key Press EN...

Страница 20: ...r Figure 1 MODBUS Protocol Data Unit MODBUS protocol mapping on a specific bus or network introduces additional fields of protocol data units The client that initiates the MODBUS exchange creates the...

Страница 21: ...s order from left to right the least significant bit LSB Maximum Significant Bit MSB Start bit 1 2 3 4 5 6 7 8 Stop bit Stop bit Figure 3 RTU pattern bit sequence Check Domain Structure Cyclic Redunda...

Страница 22: ...5 characters 1 247 Function codes conforming to MODBUS specification Data conforming to MODBUS specification CRCL CRCL No signal bytes during 3 5 characters Byte 3 5 1 n 1 1 3 5 Figure 7 MODBUS defini...

Страница 23: ...lue Low Bytes 110 0x00 Function code 0x64 Figure 10 Examples of read and hold register request and response frames MODBUS function code 0x10 write multiple registers This function code is used to writ...

Страница 24: ...mber in binary form first finding the binary representation of the integer part 17 Decimal 16 1 1 24 0 23 0 22 0 21 1 20 The binary representation of integer part 17 is 10001B then the binary represen...

Страница 25: ...0111 1011 0110 0110 0110 0110B to decimal number Step 1 Divide the binary float ing point number 0100 0010 0111 1011 0110 0110 0110 0110B into symbol bit exponential bit and mantissa bit 0 10000100 11...

Страница 26: ...hen shortcircuit the terminal Input and Ref the instrument should be displayed in 7 000 or within the error range If the display is not correct please contact the manufacturer for maintenance Please p...

Страница 27: ...e year from the date of purchase If repair is necessary and has not been the result of abuse or misuse within the warranty period please return to Icon Process Controls and amendment will be made with...

Страница 28: ...28 23 0032 01 19 23 Need a great sensor Pair your D400 Series controller with the industry s most innovative and reliable sensing technology iconprocon com...

Отзывы: