background image

 

Page 7 

6. 300B Bias Check & Adjustment  

If  you  are  unsure  about  any  aspect  of  bias  contact 
your  retailer,  Icon  Audio  or  a  competent  service 
engineer. 

 

The  Stereo  300 

uses  the  ‘Fixed  bias’  method  of  valve 

operation.  This  method  gives  higher  power,  lower 
feedback  and  cooler  running.  However  you  should 
regularly check the  bias reading  using the  built in meter 
to  ensure  best  performance  from  the  amplifier.

  This  will 

tell  you  that  the  300B  output  valves  are  operating 
normally. Adjustment is not usually needed.  This is very 
easy using the built in meter for V1 to V4 left to right. 

 

1, 

Tools you will need:

  

A small flat blade screwdriver. 

 

2, 

How to check the valves and read the meter.

 

(If changing valves see next section below) 
If possible warm up the amplifier for 10 minutes (unless 
you suspect a fault), with 

Standby “off”, volume at “zero”. 

Move the “bias” selector V1, These are the output valves 
left to right (viewed from the front). Each valve should 
have the black pointer in or near the black section. (50-
75). Precise adjustment is not necessary. This will not 
affect sound quality. 
Adjust if necessary, (see below) and repeat for V2/3/4. 
A high reading (80+) needs adjustment, as this valve is 
drawing too much current. Lower than 50 will cause no 
harm but the performance will be reduced. 

NOTE 

3. The readings are affected by changes in your local 
mains voltage. So if all four valves are slightly high or 
low this is normal. Check again another time. 

 

4. 

How to adjust the bias:

  

Make sure you are reading the valve to adjust, e.g. read 
V1 to adjust V1 screw. If the reading is incorrect, set this 
by  using  the  screw  very  slowly  up  or  down  until  the 
correct  reading  is  obtained.  They  are  very  sensitive  so 
adjust  very  carefully.  If  the  reading  appears  a  little 
unstable this is normally due to mains fluctuations. 
You  may  need  to  repeat  this  a  couple  of  times  as  the 
adjustment  of  one  valve  may  affect  the  other  readings. 
Allow 20 minutes to fully warm up. 
5, If one or more valves are showing erratic readings or 
you  cannot  set  the  correct  voltage,  then  that  valve  is 
probably  faulty  or  out  of  specification.  If  you  are  unable 
to  set  the  reading  high  enough  this  usually  means  the 
emission of the valve is too low. 

 

 

Replacing the 300B Valves 

Important:  Do  not  attempt  to  change  the  300Bs 
without  reading  these  notes.  Failure  to  do  so  could 
be both dangerous and damaging to the amplifier.  
When replacing valves, we recommend that you use 
“matched pairs” , for best performance. 
Use only 300B valves.  

 

Health & safety; High voltages are present inside the 
amplifier and on exposed valve sockets when valves 
are  removed,  so  take  suitable  care.  It  is  not 
necessary  to  remove  the  bottom  cover.  Beware 
valves get hot in operation! 

 

5,  Changing  valves:  You  should  if  possible  check  the 
bias setting before you attempt to change the valve(s), in 
order to familiarise yourself with the procedure.  

Replace  and  bias  one  valve  at  a  time.  Remove  the 
first old valve and fit the replacement. Switch on and 
measure and adjust the bias.  
Do not allow the reading to go above 95

. Don’t worry 

how low the reading goes this will not cause damage. 
Continue  in  the  same  way  and  fit  both  valves.  Do 
final  adjustment  when  the  amplifier  is  fully  warmed 
up. 
If  all  is  well  there  should  be  no  more  than  a  barely 
detectable  hum  from  the  speakers,  and  the  amplifier 
should sound OK when tested. 

 

6,  If  you  cannot  set  up  the  bias  then  the  valve  is 
probably faulty. If the valves are  brand  new, you will 
need  to  check  again  after  approximately  10  &  100 
hours, after that only occasionally or if you suspect a 
problem.  

 

7,  To  avoid  damage  to  the  amplifier  and  electric 
shock  hazard  you  must  use  only  valves  marked 
300B,  and  6SN7/6N8/CV181  Or  that  you  know  to 
be  direct  equivalents.

  Use  only  valves  which  you 

know  to  be  new  or  good  condition  and  test  the 
amplifier thoroughly before resuming normal use. 

 

8, Replacing the small valves: 

These  are  four 

6SN7/CV181  valves. 

None  of  these 

requires  any  s

et  up  procedure.  It’s  just  ‘plug  and 

play’, although care should be taken when removing 
and inserting not to break the centre spigot.  

 

Icon  Audio  are  happy  to  check  your  valves  Free  of 
Charge  if  you  send  them  to  us  in  Leicester.  We  will 
also  gladly  check  you  ST300  and  re-bias  free  of 
charge for personal callers. (By appointment please) 

 

The 3 fuses inside this amplifier are critical safety 
parts They should only be replaced with the same 
type. These are available without charge (in the UK) 
from Icon Audio. Outside the UK postage is 
chargable. 

Содержание Stereo 300 MKII

Страница 1: ...tages employing the venerable 6SN7 CV181 giving an excellent combination of the best of vintage valve sound yet accurate and dynamic to suit modern speakers and recordings In order to get the best out...

Страница 2: ...ernal wiring check Check Triode mode N A Run 24 hour test Check inputs tape monitor Output Valve Bias level mv Sound Quality Channel Balance Valve Microphony Valve Seating Hum level left right mv RF T...

Страница 3: ...ndby switch into the down position The amplifier should now be working 300Bs do not glow you can only see this by looking down vertically inside the valve With the volume control set to minimum fully...

Страница 4: ...this is personal taste Icon or your dealer will advise you As all speaker cables have losses keeping the speaker cables shorter than 10m will help You can either hard wire your cable to the amplifier...

Страница 5: ...eft in the standby mode this will save power Do not leave the amplifier switched on 24 7 Your valves will be worn out in approximately nine months Never leave the amplifier switched on when unattended...

Страница 6: ...dio 3 Hum Loop If the hum is only on one input you may have a hum loop caused by too many ground or earth connections A possible remedy for this is to use an interconnect which only has the screen con...

Страница 7: ...nough this usually means the emission of the valve is too low Replacing the 300B Valves Important Do not attempt to change the 300Bs without reading these notes Failure to do so could be both dangerou...

Страница 8: ...caps Internal wiring using PTFE silver plated cable Valves carefully matched for best performance Gold plated Input speaker terminals Inputs for CD Tape Tuner Aux iPhone Record loop with monitor swit...

Страница 9: ...ioning correctly it may be helpful to contact us first as we may be able to assist you Then we would request that you return the item to us for further action You are advised to inform us of any chang...

Страница 10: ...n 2 Lift the cardboard box off 3 Remove the bottom packing first 4 Turn the unit the right way up and remove the top packing Please be careful not to damage the small switches on the front panel When...

Отзывы: