4 - 22
ADJUSTMENT ITEM
ADJUSTMENT CONDITION
OPERATION
VALUE
WFM
CENTER METER
THRESHOULD
[WFM Center THRESH]
• Set the SG as;
Frequency
: 101.75 MHz
Level
: +4 dBu
Push [F-7] key.
(Automatic
adjustment)
WFM
S-METER
[WFM S0 LEVEL]
• Set the SG as;
Frequency
: 101.75 MHz
Level
: +4 dBu
Push [F-7] key.
(Automatic
adjustment)
[WFM S9 LEVEL]
• Set the SG as;
Level
: +34 dBu
Push [F-7] key.
[WFM S9+60 LEVEL]
• Set the SG as;
Level
: +80 dBu
Push [F-7] key.
WFM
CENTER METER
[WFM Center (±0kHz)]
• Set the SG as;
Frequency
: 101.75 MHz
Level
: +60 dBu
Push [F-7] key.
(Automatic
adjustment)
[WFM Center (-100kHz)]
(Same as above)
Push [F-7] key.
[WFM Center (+100kHz)]
(Same as above)
Push [F-7] key.
4-9 WFM ADJUSTMENT [WFM]
• Push [F-6•WFM] key to select WFM ADJUSTMENT.
• Connect the SG to [ANT1/HF ANT3] and set as;
Modulation
:
None
NOTE:
HOLD
the adjustment condition (SG level and frequency etc.) until next ADJ item is displayed.