To remove a sharp spike when setting a low sampling rate value

If the sampling rate value is too low, a sharp spike (a DC) is displayed at the LO frequency position.

The DC is a constant offset in voltage, the Left and Right (I and Q) ADCs produce, due to error.

To remove the sharp spike:

 •Set a higher sampling rate value. 

or

 •Adjust the IQ balance. (Option [F7] 

 Calibration Settings 

 DC Removal Calibration for RX)

To adjust the IQ balance

1.  Set 

RF Gain

 to “MIN” on the IC-R8600 I/Q Settings screen. (p. 3)

 • Only the sharp spike is displayed.

2.  Open “DC Removal Calibration for RX.”

3.  Click the displayed mode to set “IIR-Highpass (Auto).”

 • The software automatically adjusts the IQ balance.

4.  Set 

Mode

 to “Constant (On).”

 

L

If this setting remains “IIR-Highpass (Auto),” the software removes signals near the LO frequency.

5.  Adjust 

RF Gain

 on the IC-R8600 I/Q Settings screen.

NOTE: 

When you change the sampling rate value, adjust the IQ balance again.

a sharp spike

(a DC)

Содержание IC-R8600

Страница 1: ...mware version is 1 20 or earlier download the latest version from the Icom website and update your IC R8600 http www icom co jp world support L L See the IC R8600 Instruction manual Section 13 for det...

Страница 2: ...requirements may differ depending on your PC environment permanently running software such as antivirus software peripheral devices settings of OS and other software the operation is not guaranteed Op...

Страница 3: ...and then click Agree Click 6 Click Save as in the displayed File Download dialog 7 Select the location where you want to save the file and then click Save in the displayed File Download dialog The fi...

Страница 4: ...Control is displayed click Yes to continue 2 Select English or Japanese and then click OK L L In this document English is selected Select language Click 3 Click Next Click 4 Confirm the destination f...

Страница 5: ...anager Windows 8 1 1 Click the folder icon on the taskbar 2 Right click This PC and then click Properties 3 Click Device Manager Windows 7 1 Click the folder icon on the taskbar 2 Right click Computer...

Страница 6: ...splayed screen 1Select 2Click L L If you use the RC 28 remote encoder see page 4 L L The receiver automatically enters to the remote control mode when the HDSDR software connects to the receiver In th...

Страница 7: ...k to display the Sound Card selection screen Click to set the LO frequency Click to set the Tune frequency Main Screen Waterfall Display RF Spectrum Display RF Waterfall Display AF Spectrum Display AF...

Страница 8: ...the direct sampling system performance 6ATT Default OFF Select the attenuator setting The attenuator prevents a desired signal from becoming distorted when a very strong signal is near the frequency o...

Страница 9: ...ng the actual receiver s main dial 1 Connect the RC 28 to the PC that has HDSDR installed NOTE DO NOT connect the RC 28 to the PC with other than the supplied USB cable or through any USB hub The RC 2...

Страница 10: ...libration Settings DC Removal Calibration for RX To adjust the IQ balance 1 Set RF Gain to MIN on the IC R8600 I Q Settings screen p 3 Only the sharp spike is displayed 2 Open DC Removal Calibration f...

Отзывы: