![ICI HOTSPOT Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/ici/hotspot/hotspot_user-manual_610685012.webp)
HOTSPOT IR NON-CONTACT THERMOMETER USER MANUAL
12
11. Measuring Operations
11-1. Choosing the Temperature Unit – F1 Function
1.
Press
MODE
button for 2 seconds, screen will display: F1.
2.
Press ▼ for degrees Celsius, ▲ for degrees Fahrenheit.
11-2. Alarm Setup – F2 Function
1.
Press
MODE
button for 2 seconds, screen will display: F1.
2.
Press
MODE
again to get to F2.
3.
Press ▲ to increase the threshold by 0.1 °C (0.1 °F).
4.
Press ▼ to reduce it by 0.1 °C (0.1 °F).
THE ALARM THRESHOLD DEFAULT VALUE IS 38°C (100.4°F).
11-3. Calibration – F3 Function
To adjust the calibration of your thermometer:
1.
Press
MODE
button for 2 seconds, screen will display: F1.
2.
Press
MODE
twice to get F3.
3.
Press ▲ to increase the difference by 0.1 °C (0.1 °F).
4.
Press ▼ to reduce the difference by 0.1 °C (0.1 °F).
IN THE CASES OF SEASONAL OR ENVIRONMENTAL CHANGES
A VERIFICATION AND ADJUSTMENT SHOULD BE CARRIED OUT.
THIS FUNCTION IS ONLY EFFECTIVE IN BODY MODE.
11-4. Buzzer ON/OFF – F4 (F3) Function
1.
Press
MODE
button for 2 seconds, screen will display: F1.
2.
Press three (two) times
MODE
button to get F4 (F3).
3.
Press ▲ to open the buzzer. An icon is displayed.
4.
Press ▼ to turn off the buzzer (icon will disappear).
11-5. Exiting The Setting Mode
1.
Press